DICOM PS3.4 2026b - Service Class Specifications

KK.2.2 Operations

KK.2.2.1 Action Information

The DICOM AEs that claim conformance to this SOP Class as an SCU and/or an SCP shall support the Action Types and Action Information as specified in Table KK.2-1. Descriptions of the Attributes of the Action Information are specified in Section B.30 “Inventory Creation IOD” in PS3.3.

Table KK.2-1. Inventory Creation Requests - Action Information

Action Type Name

Action Type ID

Attribute Name

Tag

Usage (SCU/SCP)

Initiate

11

Transaction UID

(0008,1195)

1/1

Scope of Inventory Sequence

(0008,0400)

2/2

>Extended Matching Mechanisms

(0008,040F)

1C/1C

>Specific Character Set

(0008,0005)

1C/1C

>Timezone Offset from UTC

(0008,0201)

1C/1C

>Range Matching Sequence

(0008,0410)

1C/1C

>>Any DA, DT, or TM Attribute

2/2

>List of UID Matching Sequence

(0008,0411)

1C/1C

>>Any UI Attribute

1/1

>Empty Value Matching Sequence

(0008,0412)

1C/1C

>>Any Attribute

2/2

>General Matching Sequence

(0008,0413)

1C/1C

>>Any Attribute

1/1

Inventory Purpose

(0008,0401)

3/1

Inventory Level

(0008,0403)

1/1

Requested Status Interval

(0008,0414)

3/1

Request Status

12

Transaction UID

(0008,1195)

1/1

Requested Status Interval

(0008,0414)

3/1

Cancel

13

Transaction UID

(0008,1195)

1/1

Retain Instances

(0008,0415)

1/1

Pause

14

Transaction UID

(0008,1195)

1/1

Resume

15

Transaction UID

(0008,1195)

1/1


KK.2.2.1.1 Scope of Inventory Sequence

The Scope of Inventory Sequence (0008,0400) specifies constraints on Studies to be included in the Inventory. See the Attribute descriptions for the Scope of Inventory Sequence (0008,0400) in Section C.38.2.1 “Scope of Inventory Macro” in PS3.3.

The Attributes are similar to Key Attributes used for Attribute matching in the Query/Retrieve Service Class (see Section C.2.2.2). The Attribute values provided in the N-ACTION shall be matched against the values of Attributes of Studies managed by the SCP, in accordance with the requirements of this section, with matching Studies to be included in the produced Inventory. Studies that match all specified keys shall be included in the produced Inventory.

The Scope of Inventory Sequence (0008,0400) may be zero-length to indicate universal match to all Studies in the repository (i.e., a comprehensive Inventory).

KK.2.2.1.2 Inventory Level

Inventory Level (0008,0403) specifies the types of records requested for inclusion in the produced Inventory SOP Instances. Inventories may be requested with Study level records, with Study and Series level records, or with Study, Series, and Instance level records.

DICOM PS3.4 2026b - Service Class Specifications