| DICOM PS3.4 2026b - Service Class Specifications |
|---|
In the Study Root Query/Retrieve Information Model, the information is arranged into three levels that correspond to one of the three values in element (0008,0052) shown in Table C.6.2-1.
The use of the word "Images" rather than "Composite Object Instances" is historical to allow backward compatibility with previous editions of the Standard. It should not be taken to mean that Composite Object Instances of other than image type are not included at the level indicated by the value IMAGE.
The Study Root Query/Retrieve Information Model may be represented by the entity relationship diagram shown in Figure C.6-2.
Table C.6-5 defines the keys at the Study Information level of the Study Root Query/Retrieve Information Model.
A description of the Attributes of this Information Model is contained in Section C.3.
Although the Patient ID may not be globally unique, the Study Instance UID is globally unique ensuring that no two studies may be misidentified. The scope of uniqueness of the Patient ID may be specified using the Issuer of Patient ID (0010,0021).
Previously, Other Patient IDs (0010,1000) was included in this table. This Attribute has been retired. See PS3.4 2017a.
Table C.6-5. Study Level Keys for the Study Root Query/Retrieve Information Model
Attributes for the Series Level of the Study Root Query/Retrieve Information Model are the same as the Attributes for the Series Level of the Patient Root Query/Retrieve Information Model described in Section C.6.1.1.4.
Attributes for the Composite Object Instance Level of the Study Root Query/Retrieve Information Model are the same as the Attributes for the Composite Object Instance Level of the Patient Root Query/Retrieve Information Model described in Section C.6.1.1.5.
| DICOM PS3.4 2026b - Service Class Specifications |
|---|