DICOM PS3.3 2025b - Information Object Definitions |
---|
The Directory Record is based on the specification of Section F.3. It is identified by a Directory Record Type (0004,1430) of Value "WF PRESENTATION". Table F.5-49 lists the set of Keys with their associated Types for such a Directory Record Type. The description of these Keys may be found in the Modules related to the Waveform Presentation State IE of the Waveform Presentation State Information Object Definitions. This Directory Record shall be used to reference a Waveform Presentation State SOP Instance. This type of Directory Record may reference a Lower-Level Directory Entity that includes one or more Directory Records as defined in Table F.4-1.
Table F.5-49. Waveform Presentation Keys
Required if an extended or replacement character set is used in one of the Keys. |
|||
Date on which the waveform presentation was created. NoteThis date may be different from the date that the DICOM SOP Instance was created, since the presentation information contained may have been recorded earlier. |
|||
Include Table 10-12 “Content Identification Macro Attributes” |
|||
Sequence of Items where each Item includes the Attributes of one Series to which the Waveform Presentation State applies. One or more Items shall be included in this Sequence. Required if the IOD of the Waveform Presentation State SOP Instance referenced by this Directory Record includes the Waveform Presentation State Relationship Module. |
|||
Unique identifier of a Series that is part of the Study defined by the Study Instance UID (0020,000D) in the enclosing data set. |
|||
The set of SR documents containing waveform Annotations to which the Waveform Presentation State applies. These shall be of the Study defined by Study Instance UID (0020,000D) and the Series defined by Series Instance UID (0020,000E) in the enclosing Item. The referenced Instances shall have a SOP Class UID (0008,0016) of "1.2.840.10008.5.1.4.1.1.88.77" (Waveform Annotation SR Storage). One or more Items shall be included in this Sequence. Required if Structured Waveform Annotation Sequence (0040,B030) is present. |
|||
>>Include Table 10-11 “SOP Instance Reference Macro Attributes”. |
|||
The set of waveforms to which the Waveform Presentation State applies. These shall be of the Study defined by Study Instance UID (0020,000D) and the Series defined by Series Instance UID (0020,000E) in the enclosing Item. One or more Items shall be included in this Sequence. The referenced SOP Class shall be the same for all SOP Instances in a single Item of this Referenced Series Sequence (0008,1115) but may be different for different Items. NoteRequired if Referenced Instance Sequence (0008,114A) is not present. |
|||
>>Include Table 10-11 “SOP Instance Reference Macro Attributes”. |
|||
Any other Attribute of the Waveform Presentation State IE Modules |
DICOM PS3.3 2025b - Information Object Definitions |
---|