| DICOM PS3.3 2026b - Information Object Definitions |
|---|
Table C.8.27.6.3-1 specifies the Attributes of the Intravascular OCT Frame Content Functional Group.
Table C.8.27.6.3-1. Intravascular OCT Frame Content Macro Attributes
|
A Sequence that describes a general description of this image or Frame. |
|||
|
The amount to shift the Frame along the axial direction, in pixels, to correct for interferometer path-length mismatch. See Section C.8.27.6.3.1 for further explanation. |
|||
|
The A-line number (zero index) that coincides with the seam line as indicated with the First A-line Location (0052,0034). |
|||
|
The number of padded A-lines contained in the FOR PROCESSING image Frame. Required if the FOR PROCESSING image contains padded A-lines. See Section C.8.27.6.3.2 for further explanation. |
OCT Z Offset Correction (0052,0030) specifies the number of pixels to shift the A-lines of the Frame to correct for the path length difference between the OCT Sample arm and Reference arm (in air). A positive value of Z Offset correction means pixels are shifted further away from the catheter optics. A negative value of Z Offset correction means pixels are shifted closer to the catheter optics. See Figure C.8.27-2
| DICOM PS3.3 2026b - Information Object Definitions |
|---|