DICOM PS3.3 2021e - Information Object Definitions

C.8.26 Visual Field Static Perimetry Measurements Modules

C.8.26.1 Visual Field Static Perimetry Measurements Series Module

Table C.8.26.1-1 specifies the Attributes that identify and describe general information about the Visual Field Static Perimetry Measurements Series.

Table C.8.26.1-1. Visual Field Static Perimetry Measurements Series Module Attributes

Attribute Name

Tag

Type

Attribute Description

Modality

(0008,0060)

1

Type of device, process or method that originally acquired the data used to create the measurements in this Series.

Enumerated Values:

OPV

See Section C.7.3.1.1.1 for further explanation.

Referenced Performed Procedure Step Sequence

(0008,1111)

1C

Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related.

Only a single Item shall be included in this Sequence.

Required if a Performed Procedure Step SOP Class was involved in the creation of this Series.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Request Attributes Sequence

(0040,0275)

3

Sequence that contains Attributes from the Imaging Service Request.

One or more Items are permitted in this Sequence.

>Include Table 10-9 “Request Attributes Macro Attributes”

In this Module Reason for Requested Procedure Code Sequence (0040,100A) has the DCID 4256 “Visual Field Procedure Modifier”.

Note

Reason for Requested Procedure Code Sequence (0040,100A) may be used to provide whether the test in the Series was taken for screening or diagnostic (quantitative) purposes.

In this Module Scheduled Protocol Code Sequence (0040,0008) has the Defined CIDs 4250 and 4251.

Note

Scheduled Protocol Code Sequence (0040,0008) may be used to provide test pattern and test strategy.

No Baseline CIDs are defined for other Attributes.

Include Table 10-16 “Performed Procedure Step Summary Macro Attributes”

In this Module Attribute Performed Protocol Code Sequence (0040,0260) is Type 1 and the DCID 4250 “Visual Field Static Perimetry Test Patterns” and CID 4251 “Visual Field Static Perimetry Test Strategies”.

Note

Performed Protocol Code Sequence (0040,0260) is used to provide test pattern and test strategy.

In this Module Attribute Protocol Context Sequence (0040,0440) is Type 1 and the DCID 4256 “Visual Field Procedure Modifier”.

Note

Protocol Context Sequence (0040,0440) is used to provide whether the test in the Series were taken for screening or diagnostic purposes.

No Baseline CIDs are defined for other Attributes.


DICOM PS3.3 2021e - Information Object Definitions