C.8.2.3 Single-Frame CT Series Module

Table C.8.2.3-1 specifies the Attributes of the Single-Frame CT Series Module, which describe single-frame CT series.

Table C.8.2.3-1. Single-Frame CT Series Module Attributes

Attribute Name

Tag

Type

Attribute Description

Modality

(0008,0060)

1

Type of device, process or method that originally acquired the data used to create the images in this Series.

Enumerated Values:

CT

See Section C.7.3.1.1.1 for further explanation.

Presentation Intent Type

(0008,0068)

1C

Identifies the intent of the images that are contained within this Series.

Enumerated Values:

FOR PRESENTATION

FOR PROCESSING

Required if SOP Class UID (0008,0016) equals "1.2.840.10008.5.1.4.1.1.2.3" (CT Image Storage - For Processing). May be present otherwise.

See Section C.8.15.1.1.1 for further explanation.