DICOM PS3.3 2021e - Information Object Definitions

C.8.19.5 X-Ray Detector Module

Table C.8.19.5-1 contains IOD Attributes that describe an X-Ray detector.

Table C.8.19.5-1. X-Ray Detector Module Attributes

Attribute Name

Tag

Type

Attribute Description

Include Table C.8-71b “Digital X-Ray Detector Macro Attributes”

Physical Detector Size

(0018,9429)

1

Dimensions of the physical detector measured in mm as a row size followed by a column size.

Position of Isocenter Projection

(0018,9430)

1C

Position of the Isocenter projection on the detector plane measured in fractional physical detector elements as a distance along the column direction followed by a distance along the row direction from the center of the TLHC detector element of a rectangle circumscribing the physical detector area.

Required if Isocenter Reference System Sequence (0018,9462) is present.


DICOM PS3.3 2021e - Information Object Definitions