DICOM PS3.3 2024d - Information Object Definitions |
---|
This section describes the Enhanced XA/XRF Image Module. Table C.8.19.2-1 contains IOD Attributes that describe a XA/XRF Image by specializing Attributes of the General Image Module and Image Pixel Module, and adding additional Attributes.
Table C.8.19.2-1. Enhanced XA/XRF Image Module Table
Image identification characteristics. See Section C.8.19.2.1.1 for specialization. |
|||
The multiplicity of planes used simultaneously during the acquisition. See Section C.8.19.2.1.3. |
|||
Identification of the plane used to acquire this image. Note
Required if Planes in Acquisition (0018,9410) is not equal to UNDEFINED. |
|||
A number identifying the single continuous gathering of data over a period of time that resulted in this image. |
|||
The date and time that the acquisition of data that resulted in this image started. NoteThe synchronization of this time with an external clock is specified in the Synchronization Module in Acquisition Time Synchronized (0018,1800). |
|||
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. |
|||
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. See Section C.8.19.2.1.2 for specialization. |
|||
Most significant bit for pixel sample data. Each sample shall have the same high bit. Shall be one less than the value in Bits Stored (0028,0101). |
|||
Number of samples (color planes) in this image shall have a value of 1. |
|||
User defined name of the protocol used to acquire this image. |
|||
User defined description of the protocol used to acquire this image. |
|||
Identifies any acquisition technique that was used during the acquisition of the image. |
|||
Content Qualification Indicator See Section C.8.13.2.1.1 for further explanation. |
|||
Sequence that describes the orientation of the patient with respect to gravity. See Section C.8.11.5.1.2 for further explanation. Only a single Item shall be included in this Sequence. Required if Positioner Type (0018,1508) equals CARM and C-arm Positioner Tabletop Relationship (0018,9474) equals YES. May be present otherwise. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
|||
Required if needed to fully specify the orientation of the patient with respect to gravity. |
|||
Sequence that describes the orientation of the patient with respect to the head of the table. See Section C.8.4.6.1.3 for further explanation. Zero or one Item shall be included in this Sequence. Required if Positioner Type (0018,1508) equals CARM and C-arm Positioner Tabletop Relationship (0018,9474) equals YES. May be present otherwise. |
|||
>Include Table 8.8-1 “Code Sequence Macro Attributes” . |
|||
Body thickness in mm at examination location perpendicular to the table top for this Series. Note
|
|||
Indicates whether or not the image contains sufficient burned in annotation to identify the patient and date the image was acquired. |
|||
Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient. If this Attribute is absent, then the image may or may not contain recognizable visual features. |
|||
Specifies whether an Image has undergone lossy compression (at a point in its lifetime). Enumerated Values: Once this value has been set to "01" it shall not be reset. See Section C.7.6.1.1.5. |
|||
Describes the approximate lossy compression ratio(s) that have been applied to this image. |
|||
A label for the lossy compression method(s) that have been applied to this image. |
|||
The images of the corresponding plane for a Biplane acquisition device. |
|||
>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
|||
Full set of Composite SOP Instances referred to inside the Referenced Image Sequences of this SOP Instance. See Section 10.42.1 for further explanation. One or more Items shall be included in this Sequence. Required if the Referenced Image Sequence (0008,1140) is present. |
|||
>Include Table C.17-3 “Hierarchical SOP Instance Reference Macro Attributes” |
|||
Full set of Composite SOP Instances referred to inside the Source Image Sequences of this SOP Instance. See Section 10.42.1 for further explanation. One or more Items shall be included in this Sequence. Required if the Source Image Sequence (0008,2112) is present. |
|||
>Include Table C.17-3 “Hierarchical SOP Instance Reference Macro Attributes” |
|||
Non-image Composite SOP Instances that are significantly related to this Image, including waveforms that may or may not be temporally synchronized with this image. |
|||
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
Code describing the purpose of the reference to the SOP Instances. |
|||
DCID 7004 “Waveform Purpose of Reference” for referenced waveforms. |
|||
Include Table 10.41-1 “General Procedure Protocol Reference Macro Attributes” |
|||
Indicates whether or not this image is a quality control or phantom image. If this Attribute is absent, then the image may or may not be a quality control or phantom image. |
|||
See Section C.7.6.1.1.6 for further explanation. |
|||
Specifies a predefined identity transformation for the Presentation LUT such that the output of all grayscale transformations, if any, are defined to be in P-Values. |
In addition to the requirements specified in Section C.8.16.1 Image Type, the following additional requirements and Defined Terms are specified.
Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.1. No additional requirements or Defined Terms.
Value 2 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.2. No additional requirements or Defined Terms.
Value 3 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.3. No additional requirements or Defined Terms.
Value 4 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.4. The value shall be NONE.
Table C.8.19.2-2 specifies the allowed combinations of Bits Allocated (0028,0100) and Bits Stored (0028,0101).
DICOM PS3.3 2024d - Information Object Definitions |
---|