C.8.15.3.15 Metal Artifact Reduction Macro

Table C.8.15.3.15-1 specifies the Attributes of the Metal Artifact Reduction Macro.

Table C.8.15.3.15-1. Metal Artifact Reduction Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Metal Artifact Reduction Sequence

(0018,9390)

1

Contains the Attributes describing the application of metal artifact reduction (MAR).

Only a single Item shall be included in this Sequence.

>Metal Artifact Reduction Applied

(0018,9391)

1

Indicates whether or not metal artifact reduction (MAR) was applied.

Enumerated Values:

YES

NO

>Metal Artifact Reduction Algorithm Identification Sequence

(0018,9392)

3

Algorithm used for metal artifact reduction.

Only a single Item is permitted in this Sequence.

>>Include Table 10-19 “Algorithm Identification Macro Attributes”

For Algorithm Family Code Sequence (0066,002F) BCID 10036 “Metal Artifact Reduction Algorithm Family”.