Table C.8-126b specifies the Attributes of the CT Additional X-Ray Source Macro. This Sequence may be multi-valued, depending on the number of additional active X-Ray sources. It defines the X-Ray source parameters beyond the basic system.
Table C.8-126b. CT Additional X-Ray Source Macro Attributes
Contains the Attributes defining the data acquisition in a multiple X-Ray source system beyond the primary source. The primary X-Ray source is specified in the CT X-Ray Details Sequence. |
|||
The diameter in mm of the region over which data were collected. |
|||
Type of filter(s) inserted into the X-Ray beam. See Section C.8.15.3.9. |
|||
The X-Ray absorbing material used in the filter. May be multi-valued. See Annex P “Correspondence of X-Ray Filter Material Codes and Defined Terms” in PS3.16 for Defined Terms. |
|||
The exposure expressed in milliampere seconds, for example calculated from exposure time and X-Ray tube current. |
|||
The weighting factor of the data from this additional source in a multiple energy composition image. This factor incorporates the effects of Required if Required if Frame Type (0008,9007) Value 4 of this frame is ENERGY_PROP_WT. May be present otherwise. |