MR Spectroscopy FOV/Geometry Sequence
|
(0018,9103)
|
1
|
Identifies the geometry parameters of this Frame.
Only a single Item shall be included in this Sequence.
|
>Spectroscopy Acquisition Data Columns
|
(0018,9127)
|
1C
|
Number of data points in the columns direction.
Required if Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
|
>Spectroscopy Acquisition Phase Rows
|
(0018,9095)
|
1C
|
Number of Phase Encoding Rows.
Required if Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
|
>Spectroscopy Acquisition Phase Columns
|
(0018,9234)
|
1C
|
Number of Phase Encoding Columns.
Required if Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
|
>Spectroscopy Acquisition Out-of-plane Phase Steps
|
(0018,9159)
|
1C
|
Number of out-of-plane Phase Encoding steps.
Required if MR Spectroscopy Acquisition Type (0018,9200) equals PLANE and Frame Type (0008,9007) Value 1 is ORIGINAL. May be present otherwise.
|
>Percent Sampling
|
(0018,0093)
|
1C
|
Fraction of acquisition matrix lines acquired, expressed as a percent.
Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise.
|
>Percent Phase Field of View
|
(0018,0094)
|
1C
|
Ratio of field of view dimension in phase direction to field of view dimension in frequency direction, expressed as a percent.
Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise.
|