| DICOM PS3.3 2026b - Information Object Definitions |
|---|
Data Path Assignment (0028,1402) specifies the data path to use in the Enhanced Blending and Display Pipeline for each data Frames of a particular data type.
Enumerated Values:
Data Frame values are passed through the Presentation LUT to produce grayscale P-Values; no blending is performed
Data Frame values are inputs to the Primary Palette Color Lookup Table
Data Frame values are inputs to the Secondary Palette Color Lookup Table
Data Frame values having Data Path Assignment (0028,1402) of SECONDARY_HIGH are concatenated as the most significant bits with Data Frame values having Data Path Assignment of SECONDARY_LOW to form inputs to the Secondary Palette Color Lookup Table
Data Frame values having Data Path Assignment (0028,1402) of SECONDARY_LOW are concatenated as the least significant bits with Data Frame values having Data Path Assignment of SECONDARY_HIGH to form inputs to the Secondary Palette Color Lookup Table
| DICOM PS3.3 2026b - Information Object Definitions |
|---|