DICOM PS3.3 2021e - Information Object Definitions

C.7.6.13 Intervention Module

Table C.7-19 describes the Attributes of therapies (e.g., interventions during an angiographic procedure) that are associated with a Study and/or image.

Table C.7-19. Intervention Module Attributes

Attribute Name

Tag

Type

Attribute Description

Intervention Sequence

(0018,0036)

3

Sequence describing interventional therapies or procedures.

One or more Items are permitted in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 9 “Image Guided Therapeutic Procedures”.

>Intervention Status

(0018,0038)

2

Temporal relation of SOP Instance to intervention

Enumerated Values:

PRE

INTERMEDIATE

POST

NONE

>Intervention Drug Code Sequence

(0018,0029)

3

Sequence that identifies the interventional drug.

Only a single Item is permitted in this Sequence.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 10 “Interventional Drug”.

>Intervention Drug Start Time

(0018,0035)

3

Time of administration of the interventional drug.

>Intervention Drug Stop Time

(0018,0027)

3

Time of completion of administration of the intervention drug.

>Administration Route Code Sequence

(0054,0302)

3

Sequence that identifies the Administration Route.

Only a single Item is permitted in this Sequence.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 11 “Route of Administration”.

>Intervention Description

(0018,003A)

3

Further description in free form text describing the therapy or other intervention.


Note

Therapy Description (0018,0039) was included in this Module in earlier editions, but its use has been retired. See PS3.3-2003.

DICOM PS3.3 2021e - Information Object Definitions