DICOM PS3.3 2025a - Information Object Definitions |
---|
This Module defines how a display device applies waveform annotations that are stored in a separate SR document to a waveform.
Table C.39.2-1. Structured Waveform Annotation Module Attributes
Selects and provides display information for waveform annotations and measurements contained in the referenced SR document. |
|||
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
This references an SR document which contains the annotations. The Instance referenced here shall be contained in the Referenced Instance Sequence (0008,114A) in the Referenced Series Sequence (0008,1115). |
||
Selects subsets of annotations in the referenced SR document for display. If no subset is selected (i.e. this Attribute is missing or the Sequence is empty) all annotations in the referenced SR document shall be displayed. |
|||
References an annotation group number (130872, DCM, "Waveform Annotation Group Number") defined within the referenced SR document to which the display information applies. |
|||
The recommended viewing montage identified by the Montage Index (0040,B03D) in the Waveform Montage Sequence (0040,B039). |
|||
A default color triplet value used to specify the text color in which it is recommended that the text be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. |
DICOM PS3.3 2025a - Information Object Definitions |
---|