| DICOM PS3.3 2026d - Information Object Definitions |
|---|
The Matrix-based RT Imaging Geometry Macro describes the location of the imaging source and image receptor by the means of matrices with respect to the Equipment Frame of Reference.
Table C.36.2.4.2-1. Matrix-based RT Imaging Geometry Macro Attributes
|
The Value of Device Index (3010,0039) from the Acquisition Device Sequence (3002,0117) corresponding to the Acquisition Device used in this Item. Required if Value 1 of Image Type (0008,0008) has the Value ORIGINAL or the current Instance was derived from an Instance whereReferenced Defined Device Index (300A,0602) was present in the Imaging Source Position Sequence (3002,010D). May be present otherwise. |
|||
|
A rigid, homogeneous 4x4 transformation matrix that maps the Imaging Source Coordinate System to the Equipment Coordinate System. Matrix elements shall be listed in row-major order. See Section C.36.1.1.11. |
|||
|
Device-specific parameters, derived from the Device Position to Equipment Mapping Matrix (3002,010F). |
|||
|
The Value of Device Index (3010,0039) from the Acquisition Device Sequence (3002,0117) corresponding to the Acquisition Device used in this Item. Required if Value 1 of Image Type (0008,0008) has the Value ORIGINAL or the current Instance was derived from an Instance where Referenced Defined Device Index (300A,0602) was present in the Image Receptor Position Sequence (3002,010E). May be present otherwise. |
|||
|
A rigid, homogeneous 4x4 transformation matrix that maps the Image Receptor Coordinate System to the Equipment Coordinate System. Matrix elements shall be listed in row-major order. See Section C.36.1.1.12. |
|||
|
Device-specific parameters, derived from the Device Position to Equipment Mapping Matrix (3002,010F). |
|||
The Device Position to Equipment Mapping Matrix (3002,010F) is the exclusive source of information to define the location of the imaging source and the image receptor. Imaging devices will solely derive the displacement by this matrix.
Some applications, which do not act as imaging devices and cannot de-compose these matrices into device-specific parameters, may want to informatively display device-specific parameters to the user. The purpose of the Device Position Parameter Sequence (3002,0110) is to facilitate such display. These Sequences are not a substitute for the Device Position to Equipment Mapping Matrix (3002,010F).
| DICOM PS3.3 2026d - Information Object Definitions |
|---|