| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.34.14-1 specifies the Attributes of the Performed Storage Module, which contain storage parameter values for a performed imaging procedure.
Table C.34.14-1. Performed Storage Module Attributes
|
Parameter values for each storage protocol element in the Protocol. The output Instances of the referenced Acquisition or Reconstruction Protocol Elements are stored to the location specified in the Output Information Sequence. See Section C.34.13.1. One or more Items shall be included in this Sequence. The performing system might not support the requested storage protocol or not be configured for the desired destination. Note
|
|||
|
>Include Table 10.28-1 “Protocol Element Identification Macro Attributes” |
|||
|
A value corresponding to the Element Number (0018,9921) of the Acquisition Protocol Element for which data is stored. NoteSource Acquisition Protocol Element references are for storage of raw acquisition data. Reconstructed slice storage involves reference to the corresponding Reconstruction Protocol Element in the Source Reconstruction Protocol Element Number (0018,993A). This may be multi-valued if multiple acquisition protocol element data are being stored together. Required if Source Reconstruction Protocol Element Number (0018,993A) is not present. |
|||
|
The Element Number (0018,9921) corresponding to the Reconstruction Protocol Element for which data is stored. This may be multi-valued if multiple reconstruction protocol element data are being stored together. Required if Source Acquisition Protocol Element Number (0018,9938) is not present. |
|||
|
The Beam Number (300A,00C0) in the Acquisition Protocol Element (specified in Source Acquisition Protocol Element Number (0018,9938) ) for which data is stored. Multiple values may be specified if data for multiple beams is being stored.
SOP Class UID (0008,0016) has a Value of "1.2.840.10008.5.1.4.1.1.200.2" (CT Performed Procedure Protocol Storage) or "1.2.840.10008.5.1.4.1.1.200.8" (XA Performed Procedure Protocol Storage), and Source Acquisition Protocol Element Number (0018,9938) is present and only some of the beams in the Acquisition Protocol Element are to be stored. |
|||
|
Uniquely identifies the referenced SOP Class. Required if the referenced acquisition or reconstruction protocol element is not in this Instance. |
|||
|
The UID of the Protocol Instance containing the element referenced in Source Reconstruction Protocol Element Number (0018,993A) or Source Acquisition Protocol Element Number (0018,9938). Required if the referenced acquisition or reconstruction protocol element is not in this Instance. |
|||
|
References to acquired and reconstructed data objects stored as part of this storage protocol element. |
|||
|
>>Include Table 10-3c “Storage Macro Attributes” |
|||
| DICOM PS3.3 2026c - Information Object Definitions |
|---|