Visual Evaluation Test Sequence
|
(0028,7028)
|
Test Results of a visual evaluation of the display subsystem output.
May include evaluation of multiple test pattern images, as long as all Items in this Sequence are generated in a single QA test session.
One or more Items shall be included in this Sequence.
|
>Test Result
|
(0028,7029)
|
Result of a test.
Enumerated Values:
-
PASS
-
Evaluated as acceptable
-
FAIL
-
Evaluated as not acceptable
-
SKIP
-
Not evaluated
|
>Test Result Comment
|
(0028,702A)
|
User comment on the result or performance of the test.
Might include scoring values that are not intended to be machine readable.
|
>Test Pattern Code Sequence
|
(0028,702C)
|
Coded identification of the test pattern presented for visual evaluation.
Only a single Item shall be included in this Sequence.
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 8301 “Test Pattern Codes”.
|
>Referenced Image Sequence
|
(0008,1140)
|
Reference to the Image Instance containing the test pattern presented for visual evaluation.
Only a single Item shall be included in this Sequence.
Required if Test Pattern Code Sequence (0028,702C) is absent in this Item. May be present otherwise.
|
>>Test Image Validation
|
(0028,702B)
|
Indication of whether the test image was validated as matching the characteristics of the display system (e.g., resolution and bit depth).
Enumerated Values:
-
MATCHED
-
the test image was determined to match
-
UNMATCHED
-
the test image was not determined to match
|
Visual Evaluation Method Code Sequence
|
(0028,702E)
|
Coded description of the methodology used during this visual evaluation.
Only a single Item shall be included in this Sequence.
|
>Include Table 8.8-1 “Code Sequence Macro Attributes”
|
BCID 8300 “Visual Evaluation Methods”.
|