DICOM PS3.3 2021e - Information Object Definitions

C.3 Visit Modules

The following Sections specify Modules relevant to a Real World Patient visit.

C.3.1 Visit Relationship Module

Table C.3-1 defines the Attributes that reference SOP Instances related to this SOP Class.

Table C.3-1. Visit Relationship Module Attributes

Attribute Name

Tag

Attribute Description

Referenced Study Sequence

(0008,1110)

Uniquely identifies the Study SOP Instances associated with the Visit SOP Instance. One or more Items shall be included in this Sequence.

See Section 10.6.1.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced Patient Sequence

(0008,1120)

Uniquely identifies the Patient SOP Instance that relates to the Visit SOP Instance. Only a single Item shall be included in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”


DICOM PS3.3 2021e - Information Object Definitions