DICOM PS3.3 2021e - Information Object Definitions

C.13.9 Printer Module

Table C.13-9. Printer Module Attributes

Attribute Name

Tag

Attribute Description

Printer Status

(2110,0010)

Printer device status.

Enumerated Values:

NORMAL

WARNING

FAILURE

Printer Status Info

(2110,0020)

Additional information about Printer Status (2110,0010).

Defined Terms if Printer Status (2110,0010) is NORMAL:

NORMAL

See Section C.13.9.1 for Defined Terms when the Printer Status is equal to WARNING or FAILURE.

Printer Name

(2110,0030)

User defined name identifying the printer.

Manufacturer

(0008,0070)

Manufacturer of the printer.

Manufacturer Model Name

(0008,1090)

Manufacturer's model number of the printer.

Device Serial Number

(0018,1000)

Manufacturer's serial number of the printer.

Software Versions

(0018,1020)

Manufacturer's designation of software version of the printer.

Date Of Last Calibration

(0018,1200)

Date when the printer was last calibrated.

Time Of Last Calibration

(0018,1201)

Time when the printer was last calibrated.


C.13.9.1 Printer Status Info and Execution Status Info

Additional Defined Terms for Printer Status Info (2110,0020) and Execution Status Info (2100,0030) are:

Table C.13.9.1-1. Defined Terms for Printer and Execution Status Info

BAD RECEIVE MGZ

There is a problem with the film receive magazine. Films from the printer cannot be transported into the magazine.

BAD SUPPLY MGZ

There is a problem with a film supply magazine. Films from this magazine cannot be transported into the printer.

CALIBRATING

Printer is performing self calibration, it is expected to be available for normal operation shortly.

CALIBRATION ERR

An error in the printer calibration has been detected, quality of processed films may not be optimal.

CHECK CHEMISTRY

A problem with the processor chemicals has been detected, quality of processed films may not be optimal.

CHECK SORTER

There is an error in the film sorter.

CHEMICALS EMPTY

There are no processing chemicals in the processor, films will not be printed and processed until the processor is back to normal.

CHEMICALS LOW

The chemical level in the processor is low, if not corrected, it will probably shut down soon.

COVER OPEN

One or more printer or processor covers, drawers, doors are open.

ELEC CONFIG ERR

Printer configured improperly for this job.

ELEC DOWN

Printer is not operating due to some unspecified electrical hardware problem.

ELEC SW ERROR

Printer not operating for some unspecified software error.

EMPTY 8X10

The 8x10 inch film supply magazine is empty.

EMPTY 8X10 BLUE

The 8x10 inch blue film supply magazine is empty.

EMPTY 8X10 CLR

The 8x10 inch clear film supply magazine is empty.

EMPTY 8X10 PAPR

The 8x10 inch paper supply magazine is empty.

EMPTY 10X12

The 10x12 inch film supply magazine is empty.

EMPTY 10X12 BLUE

The 10x12 inch blue film supply magazine is empty.

EMPTY 10X12 CLR

The 10x12 inch clear film supply magazine is empty.

EMPTY 10X12 PAPR

The 10x12 inch paper supply magazine is empty.

EMPTY 10X14

The 10x14 inch film supply magazine is empty.

EMPTY 10X14 BLUE

The 10x14 inch blue film supply magazine is empty.

EMPTY 10X14 CLR

The 10x14 inch clear film supply magazine is empty.

EMPTY 10X14 PAPR

The 10x14 inch paper supply magazine is empty.

EMPTY 11X14

The 11x14 inch film supply magazine is empty.

EMPTY 11X14 BLUE

The 11x14 inch blue film supply magazine is empty.

EMPTY 11X14 CLR

The 11x14 inch clear film supply magazine is empty.

EMPTY 11X14 PAPR

The 11x14 inch paper supply magazine is empty.

EMPTY 14X14

The 14x14 inch film supply magazine is empty.

EMPTY 14X14 BLUE

The 14x14 inch blue film supply magazine is empty.

EMPTY 14X14 CLR

The 14x14 inch clear film supply magazine is empty.

EMPTY 14X14 PAPR

The 14x14 inch paper supply magazine is empty.

EMPTY 14X17

The 14x17 inch film supply magazine is empty.

EMPTY 14X17 BLUE

The 14x17 inch blue film supply magazine is empty.

EMPTY 14X17 CLR

The 14x17 inch clear film supply magazine is empty.

EMPTY 14X17 PAPR

The 14x17 inch paper supply magazine is empty.

EMPTY 24X24

The 24x24 cm film supply magazine is empty.

EMPTY 24X24 BLUE

The 24x24 cm blue film supply magazine is empty.

EMPTY 24X24 CLR

The 24x24 cm clear film supply magazine is empty.

EMPTY 24X24 PAPR

The 24x24 cm paper supply magazine is empty.

EMPTY 24X30

The 24x30 cm film supply magazine is empty.

EMPTY 24X30 BLUE

The 24x30 cm blue film supply magazine is empty.

EMPTY 24X30 CLR

The 24x30 cm clear film supply magazine is empty.

EMPTY 24X30 PAPR

The 24x30 cm paper supply magazine is empty.

EMPTY A4 PAPR

The A4 paper supply magazine is empty.

EMPTY A4 TRANS

The A4 transparency supply magazine is empty.

EXPOSURE FAILURE

The exposure device has failed due to some unspecified reason.

FILM JAM

A film transport error has occurred and a film is jammed in the printer or processor.

FILM TRANSP ERR

There is a malfunction with the film transport, there may or may not be a film jam.

FINISHER EMPTY

The finisher is empty.

FINISHER ERROR

The finisher is not operating due to some unspecified reason.

FINISHER LOW

The finisher is low on supplies

LOW 8X10

The 8x10 inch film supply magazine is low.

LOW 8X10 BLUE

The 8x10 inch blue film supply magazine is low.

LOW 8X10 CLR

The 8x10 inch clear film supply magazine is low.

LOW 8X10 PAPR

The 8x10 inch paper supply magazine is low.

LOW 10X12

The 10x12 inch film supply magazine is low.

LOW 10X12 BLUE

The 10x12 inch blue film supply magazine is low.

LOW 10X12 CLR

The 10x12 inch clear film supply magazine is low.

LOW 10X12 PAPR

The 10x12 inch paper supply magazine is low.

LOW 10X14

The 10x14 inch film supply magazine is low.

LOW 10X14 BLUE

The 10x14 inch blue film supply magazine is low.

LOW 10X14 CLR

The 10x14 inch clear film supply magazine is low.

LOW 10X14 PAPR

The 10x14 inch paper supply magazine is low.

LOW 11X14

The 11x14 inch film supply magazine is low.

LOW 11X14 BLUE

The 11x14 inch blue film supply magazine is low.

LOW 11X14 CLR

The 11x14 inch clear film supply magazine is low.

LOW 11X14 PAPR

The 11x14 inch paper supply magazine is low.

LOW 14X14

The 14x14 inch film supply magazine is low.

LOW 14X14 BLUE

The 14x14 inch blue film supply magazine is low.

LOW 14X14 CLR

The 14x14 inch clear film supply magazine is low.

LOW 14X14 PAPR

The 14x14 inch paper supply magazine is low.

LOW 14X17

The 14x17 inch film supply magazine is low.

LOW 14X17 BLUE

The 14x17 inch blue film supply magazine is low.

LOW 14X17 CLR

The 14x17 inch clear film supply magazine is low.

LOW 14X17 PAPR

The 14x17 inch paper supply magazine is low.

LOW 24X24

The 24x24 cm film supply magazine is low.

LOW 24X24 BLUE

The 24x24 cm blue film supply magazine is low.

LOW 24X24 CLR

The 24x24 cm clear film supply magazine is low.

LOW 24X24 PAPR

The 24x24 cm paper supply magazine is low.

LOW 24X30

The 24x30 cm film supply magazine is low.

LOW 24X30 BLUE

The 24x30 cm blue film supply magazine is low.

LOW 24X30 CLR

The 24x30 cm clear film supply magazine is low.

LOW 24X30 PAPR

The 24x30 cm paper supply magazine is low.

LOW A4 PAPR

The A4 paper supply magazine is low.

LOW A4 TRANS

The A4 transparency supply magazine is low.

NO RECEIVE MGZ

The film receive magazine not available

NO RIBBON

The ribbon cartridge needs to be replaced.

NO SUPPLY MGZ

The film supply magazine specified for this job is not available.

CHECK PRINTER

The printer is not ready at this time, operator intervention is required to make the printer available.

CHECK PROC

The processor is not ready at this time, operator intervention is required to make the printer available.

PRINTER DOWN

The printer is not operating due to some unspecified reason.

PRINTER BUSY

Printer is not available at this time, but should become ready without user intervention. This is to handle non-initialization instances.

PRINT BUFF FULL

The Printer 's buffer capacity is full. The printer is unable to accept new images in this state. The printer will correct this without user intervention. The SCU should retry later.

PRINTER INIT

The printer is not ready at this time, it is expected to become available without intervention. For example, it may be in a normal warm-up state.

PRINTER OFFLINE

The printer has been disabled by an operator or service person.

PROC DOWN

The processor is not operating due to some unspecified reason.

PROC INIT

The processor is not ready at this time, it is expected to become available without intervention. For example, it may be in a normal warm-up state.

PROC OVERFLOW FL

Processor chemicals are approaching the overflow full mark.

PROC OVERFLOW HI

Processor chemicals have reached the overflow full mark.

QUEUED

Print Job in Queue

RECEIVER FULL

The Film receive magazine is full.

REQ MED NOT INST

The requested film, paper, or other media supply magazine is installed in the printer, but may be available with operator intervention.

REQ MED NOT AVAI

The requested film, paper, or other media requested is not available on this printer.

RIBBON ERROR

There is an unspecified problem with the print ribbon.

SUPPLY EMPTY

The printer is out of film.

SUPPLY LOW

The film supply is low.

UNKNOWN

There is an unspecified problem.


DICOM PS3.3 2021e - Information Object Definitions