DICOM PS3.3 2022b - Information Object Definitions

C.11.22 Volumetric Presentation State Identification Module

Table C.11.22-1 contains Attributes that describe the volume view type of this Presentation State.

Table C.11.22-1. Volumetric Presentation State Identification Module Attributes

Attribute Name

Tag

Type

Attribute Description

Presentation Creation Date

(0070,0082)

1

Date on which this presentation was created.

Note

This date may be different from the date that the DICOM SOP Instance was created, since the Presentation State information contained may have been recorded earlier.

Presentation Creation Time

(0070,0083)

1

Time at which this presentation was created.

Note

This time may be different from the time that the DICOM SOP Instance was created, since the Presentation State information contained may have been recorded earlier.

Include Table 10-12 “Content Identification Macro Attributes”

Note

The Content Label value may be used by an application as a Defined Term in order to imply some grouping of different Presentation States, i.e., it may have the same value for different Presentation State Instances that share some common concept.

Rendered Image Reference Sequence

(0070,1104)

3

A rendered representation of the view described by this Presentation State.

Only one Item shall be present in this Sequence.

See Section XXX.2 “Volumetric Presentation States vs. Static Derived Images” in PS3.17.

>Include Table 10-3 “Image SOP Instance Reference Macro Attributes”

Presentation Display Collection UID

(0070,1101)

3

Unique identifier of a collection of Presentation State Instances that are intended by the Presentation State creator to be processed or displayed together.

See Section C.11.22.1.

Note

A Key Object Selection Document with a Document Title of (113022, DCM, "Collection of Presentation States") could be used to reference all Volumetric Presentation States in a particular spatial collection.

Presentation Sequence Collection UID

(0070,1102)

1C

Unique identifier of a collection of Presentation State Instances that are sequentially related.

See Section C.11.22.2.

Required if Presentation Animation Style (0070,1A01) is present with a value of PRESENTATION_SEQ. May be present otherwise.

Note

A Key Object Selection Document with a Document Title of (113022, DCM, "Collection of Presentation States") could be used to reference all Volumetric Presentation States in a particular Presentation Sequence Collection.

Presentation Sequence Position Index

(0070,1103)

1C

Ordinal position of this Presentation State within a Presentation Sequence Collection. Shall be a monotonically increasing integer, starting from 1, incrementing by 1, unique within the Instances sharing the same value of Presentation Sequence Collection UID (0070,1102).

See Section C.11.22.2.

Required if Presentation Sequence Collection UID (0070,1102) is present.


C.11.22.1 Presentation Display Collection UID

Views that are intended to be displayed together can be associated by assigning them to a common Presentation Display Collection UID (0070,1101).

Note

Examples of a display collection include:

  • a set of mutually-orthogonal views that are intended to be displayed together

  • a set of slices that are intended to be displayed in a "light-box" format

Display applications applying a Presentation State that contains a Presentation Display Collection UID (0070,1101) may look for other Presentation States that share the same value of Presentation Display Collection UID (0070,1101) and choose to process or present them together.

See Section XXX.3.1 “Simple Planar MPR View” in PS3.17 for an example of the use of this Attribute.

DICOM PS3.3 2022b - Information Object Definitions