DICOM PS3.3 2025a - Information Object Definitions

A.92 Waveform Presentation State Information Object Definitions

A.92.1 Waveform Presentation State IOD

A.92.1.1 Waveform Presentation State IOD Description

The Waveform Presentation State Information Object Definition (IOD) specifies information that may be used to present (display) waveforms that are referenced from within the IOD.

Note

The Waveform Presentation State object allows to store textual annotations, as well as to provide display information for annotations stored in a separate SR document. The policies related to the criteria for where specific annotations should be stored - in the Waveform Presentation State object or in the SR document - are outside the scope of the Standard.

A.92.1.2 Waveform Presentation State IOD Entity-Relationship Model

This IOD uses the E-R Model in Section A.1.2, with only the Waveform Presentation State IE below the Series IE.

A.92.1.3 Waveform Presentation State IOD Module Table

Table A.92.1-1 specifies the Modules of the Waveform Presentation State IOD.

Table A.92.1-1. Waveform Presentation State IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

Presentation Series

C.11.9

M

Frame of Reference

Synchronization

C.7.4.2

C - Required if time synchronization was applied.

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Waveform Presentation State

Presentation State Identification

C.11.10

M

Waveform Presentation State Relationship

C.39.1

M

Structured Waveform Annotation

C.39.2

U

Textual Waveform Annotation

C.39.3

U

Displayed Waveform Segment

C.39.4

U

Montage Activation

C.39.5

U

Waveform Presentation Montage

C.39.6

C - Required if Montage Activation Module is present.

SOP Common

C.12.1

M


Note

All Attributes containing color information are defined to contain PCS values, so the ICC Profile Module is not used.