A.89.4 Photoacoustic Functional Group Macros

Table A.89.4-1 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Photoacoustic IOD.

Table A.89.4-1. Photoacoustic Functional Group Macros

Functional Group Macro

Section

Usage

Frame Content

C.7.6.16.2.2

M - May not be used as a Shared Functional Group.

Pixel Measures

C.7.6.16.2.1

M

Plane Position (Patient)

C.7.6.16.2.3

C - Required if Ultrasound Acquisition Geometry (0020,9307) has a value of PATIENT. May be present otherwise. See Section A.59.4.1.2.

Plane Orientation (Patient)

C.7.6.16.2.4

C - Required if Ultrasound Acquisition Geometry (0020,9307) has a value of PATIENT. May be present otherwise. See Section A.59.4.1.2.

Referenced Image

C.7.6.16.2.5

U

Derivation Image

C.7.6.16.2.6

C - Required if the image or frame has been derived from another SOP Instance.

Frame VOI LUT

C.7.6.16.2.10

U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2.

Real World Value Mapping

C.7.6.16.2.11

U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2.

Contrast/Bolus Usage

C.7.6.16.2.12

C - Required if the Enhanced Contrast/Bolus Module is present.

Patient Orientation in Frame

C.7.6.16.2.15

U

Frame Display Shutter

C.7.6.16.2.16

U

Plane Position (Volume)

C.7.6.16.2.21

M - May not be used as a Shared Functional Group. See Section A.59.4.1.2.

Plane Orientation (Volume)

C.7.6.16.2.22

M - May not be used as a Per-Frame Functional Group. See Section A.59.4.1.2.

Temporal Position

C.7.6.16.2.23

M

Photoacoustic Excitation Characteristics

C.8.34.5.1

U

Photoacoustic Image Frame Type

C.8.34.5.2

M - May not be used as a Per-Frame Functional Group

Photoacoustic Image Data Type

C.8.34.5.3

M - May not be used as a Per-Frame Functional Group

Photoacoustic Reconstruction Algorithm Macro Attributes

C.8.34.5.4

U - May not be used as a Per-Frame Functional Group