Table A.89.4-1 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Photoacoustic IOD.
Table A.89.4-1. Photoacoustic Functional Group Macros
C - Required if Ultrasound Acquisition Geometry (0020,9307) has a value of PATIENT. May be present otherwise. See Section A.59.4.1.2. |
||
C - Required if Ultrasound Acquisition Geometry (0020,9307) has a value of PATIENT. May be present otherwise. See Section A.59.4.1.2. |
||
C - Required if the image or frame has been derived from another SOP Instance. |
||
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2. |
||
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2. |
||
C - Required if the Enhanced Contrast/Bolus Module is present. |
||
M - May not be used as a Shared Functional Group. See Section A.59.4.1.2. |
||
M - May not be used as a Per-Frame Functional Group. See Section A.59.4.1.2. |
||