Pixel Measures
|
C.7.6.16.2.1
|
M
|
Frame Content
|
C.7.6.16.2.2
|
M - May not be used as a Shared Functional Group.
|
Plane Position (Patient)
|
C.7.6.16.2.3
|
M
|
Plane Orientation (Patient)
|
C.7.6.16.2.4
|
M
|
Referenced Image
|
C.7.6.16.2.5
|
C - Required if the image or frame has been planned on another image or frame. May be present otherwise
|
Derivation Image
|
C.7.6.16.2.6
|
C - Required if the image or frame has been derived from another SOP Instance.
|
Cardiac Synchronization
|
C.7.6.16.2.7
|
C - Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
Frame Anatomy
|
C.7.6.16.2.8
|
M
|
Pixel Value Transformation
|
C.7.6.16.2.9
|
C - Required if Photometric Interpretation (0028,0004) is MONOCHROME2.
|
Frame VOI LUT
|
C.7.6.16.2.10
|
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2.
|
Real World Value Mapping
|
C.7.6.16.2.11
|
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2.
|
Contrast/Bolus Usage
|
C.7.6.16.2.12
|
C - Required if Contrast/Bolus Agent Sequence (0018,0012) is used.
|
Respiratory Synchronization
|
C.7.6.16.2.17
|
C - Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE, REALTIME or BREATH_HOLD and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Image Frame Type
|
C.8.13.5.1
|
M
|
MR Timing and Related Parameters
|
C.8.13.5.2
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR FOV/Geometry
|
C.8.13.5.3
|
C - Required if Geometry of k-Space Traversal (0018,9032) equals RECTILINEAR and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Echo
|
C.8.13.5.4
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Modifier
|
C.8.13.5.5
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Imaging Modifier
|
C.8.13.5.6
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Receive Coil
|
C.8.13.5.7
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Transmit Coil
|
C.8.13.5.8
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Diffusion
|
C.8.13.5.9
|
C - Required if Acquisition Contrast (0008,9209) in any MR Image Frame Type Functional Group in the SOP Instance equals DIFFUSION and Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Averages
|
C.8.13.5.10
|
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Spatial Saturation
|
C.8.13.5.11
|
C - Required if Spatial Pre-saturation (0018,9027) equals SLAB for any frame in the SOP Instance and Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Metabolite Map
|
C.8.13.5.12
|
C - Required if Image Type (0008,0008) Value 3 equals METABOLITE_MAP. May be present otherwise.
|
MR Velocity Encoding
|
C.8.13.5.13
|
C - Required if Phase Contrast (0018,9014) equals YES and Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.
|
MR Arterial Spin Labeling
|
C.8.13.5.14
|
C - Required if Image Type (0008,0008) Value 3 is ASL. May be present otherwise.
|
Functional MR
|
C.8.13.5.15
|
U
|
Temporal Position
|
C.7.6.16.2.23
|
U
|