DICOM PS3.3 2021e - Information Object Definitions

A.33.3 Pseudo-color Softcopy Presentation State IOD

A.33.3.1 Pseudo-color Softcopy Presentation State IOD Description

The Pseudo-Color Softcopy Presentation State Information Object Definition (IOD) specifies information that may be used to present (display) images that are referenced from within the IOD.

It includes capabilities for specifying:

  1. the output color space in PCS-Values

  2. grayscale contrast transformations including modality and VOI LUT

  3. a color palette to map the transformed grayscale values into pseudo-color

  4. mask subtraction for multi-frame images

  5. selection of the area of the image to display and whether to rotate or flip it

  6. image and display relative annotations, including graphics, text and overlays

Note

This IOD does not support presentation control for a set of images for which the modality LUT varies on a frame-by-frame basis. Other Presentation State IODs (standard or private) may support such images.

DICOM PS3.3 2021e - Information Object Definitions