A.29.3 RT Beams Treatment Record IOD Module Table

Table A.29.3-1 specifies the Modules of the RT Beams Treatment Record IOD.

Table A.29.3-1. RT Beams Treatment Record IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

RT Series

C.8.8.1

M

Clinical Trial Series

C.7.3.2

U

Equipment

General Equipment

C.7.5.1

M

Treatment Record

RT General Treatment Record

C.8.8.17

M

RT Patient Setup

C.8.8.12

U

RT Treatment Machine Record

C.8.8.18

M

Measured Dose Reference Record

C.8.8.19

U

Calculated Dose Reference Record

C.8.8.20

U

RT Beams Session Record

C.8.8.21

C - Required if Treatment Record Content Origin (300A,0709) is absent, or is present and has Value DEVICE or SIMULATION

RT Beams Salvage Record

C.8.8.31

C – Required if Treatment Record Content Origin (300A,0709) is present and has Value USER

RT Treatment Summary Record

C.8.8.23

U

General Reference

C.12.4

U

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U


Note

The Curve Module (Retired) was previously included in this IOD but has been retired. See PS3.3-2004.