DICOM PS3.3 2022b - Information Object Definitions

A.27.3 Digital Mammography X-Ray Image IOD Module Table

Table A.27-1 specifies the Modules of the Digital Mammography X-Ray Image IOD.

Table A.27-1. Digital Mammography X-Ray Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

DX Series

C.8.11.1

M

Mammography Series

C.8.11.6

M

Frame of Reference

Frame of Reference

C.7.4.1

C - Required if multiple images are obtained without releasing breast compression

Equipment

General Equipment

C.7.5.1

M

Acquisition

General Acquisition

C.7.10.1

M

Image

General Image

C.7.6.1

M

General Reference

C.12.4

U

Image Pixel

C.7.6.3

M

Contrast/Bolus

C.7.6.4

U

Display Shutter

C.7.6.11

U

Device

C.7.6.12

U

Intervention

C.7.6.13

U

Specimen

C.7.6.22

U

DX Anatomy Imaged

C.8.11.2

M

DX Image

C.8.11.3

M

DX Detector

C.8.11.4

M

X-Ray Collimator

C.8.7.3

U

DX Positioning

C.8.11.5

U

X-Ray Tomography Acquisition

C.8.7.7

U

X-Ray Acquisition Dose

C.8.7.8

U

X-Ray Generation

C.8.7.9

U

X-Ray Filtration

C.8.7.10

U

X-Ray Grid

C.8.7.11

U

Mammography Image

C.8.11.7

M

Overlay Plane

C.9.2

C - Required if graphic annotation is present - See Section A.27.3

VOI LUT

C.11.2

C - Required if Presentation Intent Type (0008,0068) is FOR PRESENTATION. Shall not be present otherwise.

Image Histogram

C.11.5

U

Acquisition Context

C.7.6.14

M

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U


Note

  1. The Curve Module (Retired) was previously included in this IOD but has been retired. See PS3.3-2004.

  2. The Specimen Identification Module was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS3.3-2008.

DICOM PS3.3 2022b - Information Object Definitions