DICOM PS3.3 2024b - Information Object Definitions

A.14.3 XA Image IOD Module Table

Table A.14-1 specifies the Modules of the XA Image IOD.

Table A.14-1. X-Ray Angiographic Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

Frame of Reference

Synchronization

C.7.4.2

U

Equipment

General Equipment

C.7.5.1

M

Acquisition

General Acquisition

C.7.10.1

M

Image

General Image

C.7.6.1

M

General Reference

C.12.4

U

Image Pixel

C.7.6.3

M

Contrast/Bolus

C.7.6.4

C - Required if contrast media was used in this Image

Cine

C.7.6.5

C - Required if pixel data is Multi-frame Cine data

Multi-frame

C.7.6.6

C - Required if pixel data is Multi-frame Cine data

Frame Pointers

C.7.6.9

U

Mask

C.7.6.10

C - Required if the Image may be subtracted

Display Shutter

C.7.6.11

U

Device

C.7.6.12

U

Intervention

C.7.6.13

U

Specimen

C.7.6.22

U

X-Ray Image

C.8.7.1

M

X-Ray Acquisition

C.8.7.2

M

X-Ray Collimator

C.8.7.3

U

X-Ray Table

C.8.7.4

C - Required if Image is created with table motion, may be present otherwise

XA Positioner

C.8.7.5

M

DX Detector

C.8.11.4

U

Overlay Plane

C.9.2

U

Multi-frame Overlay

C.9.3

C - Required if Overlay data contains multiple frames.

Modality LUT

C.11.1

C - Required if Pixel Intensity Relationship (0028,1040) is LOG

U - Optional if Pixel Intensity Relationship (0028,1040) is DISP

VOI LUT

C.11.2

U

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


Note

The Curve Module (Retired) was previously included in the Image IE for this IOD but has been retired. See PS3.3-2004.

DICOM PS3.3 2024b - Information Object Definitions