| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8-45. RT General Plan Module Attributes
|
To make assertions about Approval Status (300E,0002), use Assertion Code Sequence (0044,0101) BCID 9584 “RT Plan Approval Assertion”. The Approver is recorded in the Asserter Identification Sequence inside the Assertion Macro. |
|||
|
Describes whether RT Plan is based on patient or treatment device geometry. See Section C.8.8.9.1. |
|||
|
The RT Structure Set on which the RT Plan is based. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Related Instances of RT Dose (for grids and named/unnamed point doses). One or more Items are permitted in this Sequence. See Note 1. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Relationship of referenced plan with respect to current plan. Defined Terms:
|
|||
|
Frame of Reference to Displayed Coordinate System Transformation Matrix |
A 4x4 transformation matrix that transforms a coordinate of the Frame of Reference to a displayed coordinate system. Only rigid transformation matrices are permitted (see definition in Section C.20.2.1.2). |
||
An RT Dose IOD referenced within the Referenced Dose Sequence (300C,0080) can be used for storing grid-based (pixel) data, individual dose points (with optional dose point names), isodose curves, and DVHs.
An RT Plan Geometry (300A,000C) of PATIENT shall signify that an RT Structure Set has been defined upon which the plan geometry is based, and this RT Structure Set shall be specified in the Referenced Structure Set Sequence (300C,0060). An RT Plan Geometry (300A,000C) of TREATMENT_DEVICE shall indicate that no patient geometry is available, and that the RT Plan is being defined with respect to the IEC FIXED Coordinate System.
| DICOM PS3.3 2026d - Information Object Definitions |
|---|