DICOM PS3.3 2026d - Information Object Definitions

C.8.33.3 Summary Statistics Macro

This Macro encodes summary statistics derived from a set of values.

Table C.8.33-3. Table Summary Statistics Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Concept Name Code Sequence

(0040,A043)

1

The value (quantity) for which the statistic is a summary.

Only a single Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

CID defined by invocation

Modifier Code Sequence

(0040,A195)

1

Specified the statistic and modifies the Concept Name Code Sequence (0040,A0143) of this Macro.

Only a single Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

DCID 7464 “General Region of Interest Measurement Modifier”

Measurement Units Code Sequence

(0040,08EA)

1

Units of measurement for the statistic.

Only a single Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

DCID 82 “Measurement Unit”.

Floating Point Value

(0040,A161)

1

The value of the statistic.

Only a single Value shall be present.


DICOM PS3.3 2026d - Information Object Definitions