| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8.31.4-1 specifies the Attributes of the Breast X-Ray Geometry Macro.
Table C.8.31.4-1. Breast X-Ray Geometry Macro Attributes
|
Sequence containing the geometric properties for this Frame or set of Frames. |
|||
|
Distance in mm from source to detector center on the chest wall line. Required if Presentation Intent Type (0008,0068) = FOR PROCESSING. May be present otherwise. Note
|
|||
|
Distance in mm from source to the Breast Support side that is closest to the Imaging Subject, as measured along the X-Ray beam vector. Required if Presentation Intent Type (0008,0068) = FOR PROCESSING. May be present otherwise. Note
|
|||
|
Distance from source to isocenter in mm. Required if Presentation Intent Type (0008,0068) = FOR PROCESSING. May be present otherwise. |
|||
|
Distance in mm from the source to the surface of the patient closest to the source during the acquisition of this image. |
|||
|
Ratio of Source Image Distance (SID) over Source Object Distance (SOD). |
| DICOM PS3.3 2026d - Information Object Definitions |
|---|