| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8.31.2-1 specifies the Attributes of the Breast X-Ray Positioner Macro.
Table C.8.31.2-1. Breast X-Ray Positioner Macro Attributes
|
A Sequence that describes the geometrical position of the positioner. |
|||
|
Signed position in degrees of the X-Ray beam vector in the coronal anatomical plane as if the patient were standing facing the equipment where vertical is zero. Required if X-Ray source can be moved in the coronal anatomical plane. May be present otherwise. |
|||
|
The interpretation of the sign of the Positioner Primary Angle (0018,1510). Enumerated Values:
See Figure C.8.31.2-1. Required if Positioner Primary Angle (0018,1510) is present. |
|||
|
Position in degrees of the X-Ray beam vector in the sagittal anatomical plane as if the patient were standing where movement of the X-Ray source from anterior to posterior is positive, and vertical is zero. Required if the X-Ray source can be moved in the sagittal anatomical plane. May be present otherwise. |
Figure C.8.31.2-1 illustrates the Defined Terms for Positioner Primary Angle Direction (0018,9559), indicating which C-arm movement direction defines a positive value for Positioner Primary Angle (0018,1510).
| DICOM PS3.3 2026d - Information Object Definitions |
|---|