| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8.30.2-1 specifies the Attributes of the Corneal Topography Map Image Module, which describe an Image produced by corneal topography mapping devices.
Table C.8.30.2-1. Corneal Topography Map Image Module Attributes
|
The date and time that the acquisition of data that resulted in this image started. NoteThe synchronization of this time with an external clock is specified in the Synchronization Module in Acquisition Time Synchronized (0018,1800). |
|||
|
Image identification characteristics. See Section C.8.30.2.1.1 for specialization. |
|||
|
Number of samples (planes) in this image. See Section C.7.6.3.1.1 for further explanation. |
|||
|
Specified the intended interpretation of the pixel data. See Section C.7.6.3.1.2 for further explanation. |
|||
|
Include Table 10-10 “Basic Pixel Spacing Calibration Macro Attributes” |
|||
|
Ratio of the vertical size and horizontal size of the pixels in the image specified by a pair of integer values where the first Value refers to the vertical extent of a pixel, and the second Value refers to the horizontal extent of a pixel. See Section C.7.6.3.1.7. |
|||
|
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. See PS3.5 for further explanation. |
|||
|
Number of bits stored for each pixel sample. Shall be equal to Bits Allocated (0028,0100). |
|||
|
Most significant bit for pixel sample data. Each sample shall have the same high bit. High Bit (0028,0102) shall be one less than Bits Stored (0028,0101). |
|||
|
Specifies whether an Image has undergone lossy compression. Enumerated Values: See Section C.7.6.1.1.5 for further explanation. |
|||
|
Describes the approximate lossy compression ratio(s) that have been applied to this image. |
|||
|
A label for the lossy compression method(s) that have been applied to this image. |
|||
|
Indicates that the image does not contain burned in annotation to identify the patient and date the image was acquired. |
|||
|
Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient. |
|||
|
Laterality of object imaged (as described in Anatomic Region Sequence (0008,2218)) examined. NoteThis Attribute is mandatory, in order to ensure that images may be positioned correctly relative to one another for display. Shall be consistent with any laterality information contained in Primary Anatomic Structure Modifier Sequence (0008,2230), if present. |
|||
|
Include Table C.7.6.16-12 “Real World Value Mapping Macro Attributes” |
DCID 4267 “Corneal Topography Mapping Unit for Real World Value Mapping” |
||
|
Include Table 10-5 “General Anatomy Mandatory Macro Attributes” |
The concept code for Anatomic Region Sequence (0008,2218) shall be (81745001, SCT, "Eye"), and DCID 244 “Laterality” shall be used for Anatomic Region Modifier Sequence (0008,2220). DCID 4266 “Ophthalmic Anatomic Structure Reference Point” shall be used for Primary Anatomic Structure Sequence (0008,2228). Only a single Item shall be included in this Sequence. |
||
The Corneal Topography Map SOP Class supports a sequence of transformations that completely define the conversion of a stored image into a displayed image.
The sequence of transformations from corneal measurements to a displayable image is explicitly defined in Figure C.8.30.2.1-1. Figure C.8.30.2.1-1 also conveys the transformation to display the scaling of the color map using its Section C.7.6.16.2.11 Real World Value Mapping Macro.
Requirements for an SCP of the Corneal Topography Map Storage SOP Class when rendering SOP Instances of this SOP Class are defined in Section B.5.1.17 “Corneal Topography Map Storage SOP Class” in PS3.4. Formerly, these requirements could be found in this Section of PS3.3.
| DICOM PS3.3 2026d - Information Object Definitions |
|---|