| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8.27.4-1 specifies the Attributes of the Intravascular OCT Processing Parameters Module, which describe the parameters used for processing an Intravascular Optical Coherence Tomography polar data image.
Table C.8.27.4-1. Intravascular OCT Processing Parameters Module Attributes
|
Indicates whether or not Z offset correction was applied to the OCT image Frames. |
|||
|
Indicates whether or not Refractive Index correction was applied. See Section C.8.27.4.1 for further explanation. |
|||
|
The relationship between the pixel sample values and the IVOCT A-line intensity. See Section C.8.27.4.2. |
|||
|
Orientation in degrees at which the 1st A-line in the Frame should be displayed. Value shall be between 0 and 360, with zero representing vertical. |
For Refractive Index Applied (0052,003A) a Value of YES indicates that Effective Refractive Index (0052,0004) correction has been applied to the reported A-line Pixel Spacing (0052,0014). A Value of NO indicates that Effective Refractive Index (0052,0004) correction has not been applied (i.e., A-line Pixel Spacing corresponds to the pixel spacing in air).
| DICOM PS3.3 2026c - Information Object Definitions |
|---|