| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8.25.7-1 specifies the Attributes of the General Ophthalmic Refractive Measurements Module.
Table C.8.25.7-1. General Ophthalmic Refractive Measurements Module Attributes
|
A Sequence that specifies Ophthalmic Refractive Measurements SOP Instances that are relevant to the interpretation of this SOP Instance. Zero or more Items shall be included in this Sequence. See Section C.8.25.7.1.1 for further explanation. Required if Visual Acuity Type Code Sequence (0046,0121) is present. May be present otherwise. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
Whenever visual acuity is measured it is necessary to know the conditions under which it was measured. If the vision was measured with the patient's habitual correction in place, this Attribute is used to reference a lensometry measurements SOP Instance for the current glasses. If the measurements are of the patient's best-corrected vision, this Attribute may reference the SOP Instance that describes the correction that resulted in that vision, which may for example be the subjective refraction, or the autorefraction.
This Attribute enables a receiver of this SOP Instance to identify the results of measurements from a lensometer, autorefractor, or automated phoropter (subjective refraction) that are relevant to interpretation.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|