DICOM PS3.3 2026d - Information Object Definitions
C.8.25.14.2 Ophthalmic Axial Length Segmental Measurements Macro

Table C.8.25.14-3 specifies the Attributes of the Ophthalmic Axial Length Segmental Measurements Macro.

Table C.8.25.14-3. Ophthalmic Axial Length Segmental Measurements Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Ophthalmic Axial Length

(0022,1019)

1

The axial length measurement, in mm.

The type of measurement is specified in the Ophthalmic Axial Length Measurements Type (0022,1010).

See Section C.8.25.14.1.1.3 and Section C.8.25.14.1.1.4 for further explanation.

Ophthalmic Axial Length Measurement Modified

(0022,1140)

1

Whether or not the clinician intervened to modify the output of the device. For example by forcing it to select a different peak in the display.

Enumerated Values:

YES

NO

Ophthalmic Axial Length Measurements Segment Name Code Sequence

(0022,1101)

1

The name of the segment measured. See Section C.8.25.14.1.1.4 for further explanation.

Only a single Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

DCID 4233 “Ophthalmic Axial Length Measurements Segment Name”

Include Table C.8.25.14-4 “Ophthalmic Axial Measurements Related Information Macro Attributes”


DICOM PS3.3 2026d - Information Object Definitions