| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8.25.10-1 defines the Attributes used to represent the keratometric measurements by a keratometer of a patient's eyes. Usually both eyes are measured, but sometimes just one.
Table C.8.25.10-1. Keratometry Measurements Module Attributes
|
A Sequence that specifies keratometric measurements of a patient's right eye, defining principal meridians wherein the steepest meridian is separated by 90 degrees from the flattest. Only a single Item shall be included in this Sequence. Required if the right eye is measured. Note
|
|||
|
>Include Table C.8.25.10-2 “Keratometric Measurements Macro Attributes” |
|||
|
A Sequence that specifies keratometric measurements of a patient's left eye, defining principal meridians wherein the steepest meridian is separated by 90 degrees from the flattest. Only a single Item shall be included in this Sequence. Required if the left eye is measured. Note
|
|||
|
>Include Table C.8.25.10-2 “Keratometric Measurements Macro Attributes” |
|||
Table C.8.25.10-2 specifies the Attributes of the Keratometric Measurements Macro.
Table C.8.25.10-2. Keratometric Measurements Macro Attributes
| DICOM PS3.3 2026c - Information Object Definitions |
|---|