DICOM PS3.3 2026b - Information Object Definitions

C.8.24.4 IVUS Image Module

Table C.8.24.4-1 specifies the Attributes of the IVUS Image Module.

Table C.8.24.4-1. IVUS Image Module Attributes

Attribute Name

Tag

Type

Attribute Description

IVUS Acquisition

(0018,3100)

1

Defined Terms:

MOTOR_PULLBACK

MANUAL_PULLBACK

SELECTIVE

GATED_PULLBACK

See Section C.8.5.6.1.21

IVUS Pullback Rate

(0018,3101)

1C

Required if IVUS Acquisition (0018,3100) Value is MOTOR_PULLBACK. Specified in units of mm/sec.

See Section C.8.5.6.1.22

IVUS Gated Rate

(0018,3102)

1C

Required if IVUS Acquisition (0018,3100) Value is GATED_PULLBACK. Specified in units of mm/beat.

See Section C.8.5.6.1.23

IVUS Pullback Start Frame Number

(0018,3103)

1C

Required if IVUS Acquisition (0018,3100) Value is MOTOR_PULLBACK or GATED_PULLBACK.

See Section C.8.5.6.1.24

IVUS Pullback Stop Frame Number

(0018,3104)

1C

Required if IVUS Acquisition (0018,3100) Value is MOTOR_PULLBACK or GATED_PULLBACK.

See Section C.8.5.6.1.25


DICOM PS3.3 2026b - Information Object Definitions