| DICOM PS3.3 2026b - Information Object Definitions |
|---|
Table C.8.21.2.3-1 specifies the Attributes of the Breast Tomosynthesis Contributing Sources Module, which describe Breast Tomosynthesis Contributing Sources. The Attributes in this Module specify the overall characteristics of one or more source images that were used to create a Breast Tomosynthesis Image SOP Instance.
The intention of this Module is to supply acquisition information without the need to have access to all the source projection images.
Table C.8.21.2.3-1. Breast Tomosynthesis Contributing Sources Module Attributes
|
A Sequence that describes characteristics of the sources that are used to create a derived SOP Instance. |
|||
|
>Include Table 10-13 “General Contributing Sources Macro Attributes” |
|||
|
>Include Table 10-14 “Contributing Image Sources Macro Attributes” |
|||
|
The ID or serial number of the detector used to acquire this image. NoteDue to the length limit imposed by the VR of this Attribute, a full-length serial number may need to be truncated. If truncated, the remaining string value should be as unique as possible within the local (site) context of use, and hence perhaps retain the tail rather than the head of the string. The full-length value can be encoded in X-Ray Detector ID (0018,9371). |
|||
|
Identifier of the physical X-Ray detector. This might be the serial number. |
|||
|
The date on which the detector used to acquire this image was last calibrated. |
|||
|
The time at which the detector used to acquire this image was last calibrated. |
|||
|
Physical distance between the center of each detector element, specified by a numeric pair: row spacing value (delimiter) column spacing value in mm. See Section 10.7.1.3 for further explanation of the order of the Values. |
|||
| DICOM PS3.3 2026b - Information Object Definitions |
|---|