DICOM PS3.3 2026c - Information Object Definitions

C.8.20 Segmentation

This section describes the specific Modules for the Segmentation IOD.

C.8.20.1 Segmentation Series Module

Table C.8.20-1 specifies the Attributes of the Segmentation Series Module.

Table C.8.20-1. Segmentation Series Module Attributes

Attribute Name

Tag

Type

Attribute Description

Modality

(0008,0060)

1

Type of device, process or method that created the Instances in this Series

Enumerated Values:

SEG

Series Number

(0020,0011)

1

A number that identifies this Series.

Referenced Performed Procedure Step Sequence

(0008,1111)

1C

Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related

Only a single Item shall be included in this Sequence.

Required if a Performed Procedure Step SOP Class was involved in the creation of this Series.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”


DICOM PS3.3 2026c - Information Object Definitions