DICOM PS3.3 2026d - Information Object Definitions

C.8.20.2 Segmentation Image Module

Table C.8.20-2 specifies the Attributes of the Segmentation Image Module.

Table C.8.20-2. Segmentation Image Module Attributes

Attribute Name

Tag

Type

Attribute Description

Image Type

(0008,0008)

1

Image identification characteristics.

Value 1 shall be DERIVED. Value 2 shall be PRIMARY. No other values shall be present.

Include Table 10-12 “Content Identification Macro Attributes”

Samples per Pixel

(0028,0002)

1

Number of samples (planes) in this image.

Enumerated Values:

1

Photometric Interpretation

(0028,0004)

1

Specifies the intended interpretation of the pixel data.

Enumerated Values if Segmentation Type (0062,0001) is BINARY or FRACTIONAL:

MONOCHROME2

Enumerated Values if Segmentation Type (0062,0001) is LABELMAP:

MONOCHROME2

PALETTE COLOR

Pixel Representation

(0028,0103)

1

Data representation of pixel samples.

Enumerated Values:

0

Bits Allocated

(0028,0100)

1

Number of bits allocated for each pixel sample.

See Section C.8.20.2.1.

Enumerated Values if Segmentation Type (0062,0001) is BINARY:

1

Enumerated Values if Segmentation Type (0062,0001) is FRACTIONAL:

8

Enumerated Values if Segmentation Type (0062,0001) is LABELMAP:

8

16

Bits Stored

(0028,0101)

1

Number of bits stored for each pixel sample.

See Section C.8.20.2.1.

Enumerated Values if Segmentation Type (0062,0001) is BINARY:

1

Enumerated Values if Segmentation Type (0062,0001) is FRACTIONAL:

8

Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 8:

8

Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 16:

16

High Bit

(0028,0102)

1

Most significant bit for pixel sample data.

See Section C.8.20.2.1.

Enumerated Values if Segmentation Type (0062,0001) is BINARY:

0

Enumerated Values if Segmentation Type (0062,0001) is FRACTIONAL:

7

Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 8:

7

Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 16:

15

Lossy Image Compression

(0028,2110)

1

Specifies whether an Image has undergone lossy compression (at a point in its lifetime), or is derived from lossy compressed images.

Enumerated Values:

00

Image has not been subjected to lossy compression.

01

Image has been subjected to lossy compression.

Once this Attribute has been set to a Value of "01" it shall not be reset.

See Section C.8.20.2.2 and Section C.7.6.1.1.5, “Lossy Image Compression”.

Lossy Image Compression Ratio

(0028,2112)

1C

Describes the approximate lossy compression ratio(s) that have been applied to this image.

See Section C.7.6.1.1.5.2.

Required if present in the source images or this IOD Instance has been compressed.

Lossy Image Compression Method

(0028,2114)

1C

A label for the lossy compression method(s) that have been applied to this image.

See Section C.7.6.1.1.5.1.

Required if present in the source images or this IOD Instance has been compressed. See Section C.8.20.2.2.

Segmentation Type

(0062,0001)

1

The type of encoding used to indicate the presence of the segmented property at a pixel/voxel location.

Enumerated Values:

BINARY

FRACTIONAL

LABELMAP

See Section C.8.20.2.3.

Segmentation Fractional Type

(0062,0010)

1C

For fractional segmentation encoding, the meaning of the fractional value.

Required if Segmentation Type (0062,0001) is FRACTIONAL.

See Section C.8.20.2.3 for Enumerated Values.

Maximum Fractional Value

(0062,000E)

1C

Specifies the value that represents a probability of 1 or complete occupancy. There shall be no values in Pixel Data (7FE0,0010) greater than this value. Required if Segmentation Type (0062,0001) is FRACTIONAL.

Segments Overlap

(0062,0013)

3

Whether or not any Segments in this Instance overlap. I.e., whether or not any pixel is or might be in more than one Segment.

Enumerated Values:

YES

Some Segments overlap

UNDEFINED

Some Segments might overlap

NO

No Segments overlap

See Section C.8.20.2.3.

If present, shall be NO if Segmentation Type (0062,0001) is LABELMAP.

Note

If the value is NO, then a receiving application to which this matters can be assured that no Segments overlap and does not need to check every pixel. If the value is UNDEFINED or YES, or the Attribute is absent, then a receiving application might need to check every pixel in every Segment.

Segment Sequence

(0062,0002)

1

Describes the Segments that are contained within the data.

One or more Items shall be included in this Sequence.

Note

The Items of this Sequence are not required to be in any particular order, i.e., are not required to be ordered by Segment Number (0062,0004).

>Include Table C.8.20-4 “Segment Description Macro Attributes”

>Segment Algorithm Name

(0062,0009)

1C

The name(s) of algorithm(s) used to generate the Segment. Required if Segment Algorithm Type (0062,0008) is not MANUAL.

>Segmentation Algorithm Identification Sequence

(0062,0007)

3

A description of how this Segment was derived.

Algorithm Name (0066,0036) within this Sequence may be identical to Segment Algorithm Name (0062,0009).

One or more Items are permitted in this Sequence.

Note

Previously, the Segment Surface Generation Algorithm Identification Code Sequence (0066,002D) was used, but it has been replaced in this Module, since not all segmentation algorithms involve surface generation. See PS3.3-2016d.

>>Include Table 10-19 “Algorithm Identification Macro Attributes”

BCID 7162 “Surface Processing Algorithm Family”.

>Recommended Display Grayscale Value

(0062,000C)

3

A default single gray unsigned value in which it is recommended that the maximum pixel value in this Segment be rendered on a monochrome display. The units are specified in P-Values from a minimum of 0000H (black) up to a maximum of FFFFH (white).

Note

The maximum P-Value for this Attribute may be different from the maximum P-Value from the output of the Presentation LUT, which may be less than 16 bits in depth.

>Recommended Display CIELab Value

(0062,000D)

3

A default color value in which it is recommended that Segment be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1.

Shall not be present if Segmentation Type (0062,0001) is LABELMAP and Photometric Interpretation (0028,0004) is PALETTE COLOR.


C.8.20.2.1 Bits Allocated, Bits Stored, and High Bit

As a consequence of the Enumerated Values for Bits Allocated, Bits Stored, and High Bit Attributes, single bit pixels shall be packed 8 to a byte as defined by the encoding rules in PS3.5.

DICOM PS3.3 2026d - Information Object Definitions