| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8.19.6-4 specifies the Attributes of the X-Ray Frame Pixel Data Properties Macro.
Table C.8.19.6-4. X-Ray Frame Pixel Data Properties Macro Attributes
|
Sequence containing the pixel data properties for this Frame. |
|||
|
Type of Frame. A multi-valued Attribute analogous to Image Type (0008,0008). Enumerated Values and Defined Terms are the same as those for the Values of Image Type (0008,0008). See Section C.8.19.2.1.1. |
|||
|
The relationship between the Pixel and the X-Ray beam intensity. See Section C.8.19.6.4.1.1. |
|||
|
The sign of the relationship between the pixel sample values stored in Pixel Data (7FE0,0010) and the X-Ray beam intensity. Enumerated Values: See Section C.8.11.3.1.2 for further explanation. |
|||
|
Physical distance measured at the receptor plane of the detector between the centers of each pixel specified by a numeric pair - row spacing value (delimiter) column spacing value in mm. See Section 10.7.1.3 for further explanation of the order of the Values. The Value of this Attribute shall never be adjusted to account for calibration against an object of known size; Pixel Spacing (0028,0030) is specified for that purpose. It is only allowed to be adjusted to compensate for the change of the Field of View Dimension(s) in Float (0018,9461). See Section C.8.19.6.4.1.2 Required if Image Type (0008,0008) Value 1 equals ORIGINAL. May be present otherwise. |
|||
|
Offset of the TLHC of the image stored in Pixel Data (7FE0,0010) from the TLHC of the Field Of View Area after FOV rotation and Flip. It is measured in image stored pixels as a row offset followed by a column offset. |
|||
|
Angle clockwise of the row direction of the image stored in Pixel Data (7FE0,0010) relative to the row direction of the Field Of View Area after FOV rotation and Flip. It is measured in degrees. |
|||
|
Geometrical characteristics of the pixel data to indicate whether pixel spacing is uniform for all pixels or not. |
|||
|
The percentage of the maximum deviation of the pixel spacing values of images for which the geometric properties are non-uniform. Required if Geometrical Properties (0028,9444) equals NON_UNIFORM. |
|||
|
The type or a combination of types of image processing applied to the pixel data before being stored. |
Pixel Intensity Relationship (0028,1040) shall identify the relationship of the pixel values to the X-Ray beam intensity.
Defined Terms:
Non-linear "Log Function"; A Pixel Intensity Relationship LUT shall be included with the image to allow it to be mapped back to its proportional value to X-Ray beam intensity.
Not proportional to X-Ray beam intensity. If a TO_LINEAR Pixel Intensity Relationship LUT Item is supplied, scaling back to X-Ray beam intensity is possible.
The two Values of the Imager Pixel Spacing (0018,1164) are related to the value(s) of the Field of View Dimension(s) in Float (0018,9461) in the X-Ray Field of View Functional Group of this Frame and the Values of the Rows (0028,0010) and Columns (0028,0011).
The value(s) of Field of View Dimension(s) in Float (0018,9461) may change on a Frame by Frame base. The values of the Rows and Columns Attributes are equal for all Frames. The relationship between the Attributes depends on the Field of View Shape (0018,1147):
If the Field of View Shape (0018,1147) equals RECTANGLE:
Imager Pixel Spacingrow spacing = FOV Dimensionrow dimension / Rows
Imager Pixel Spacingcolumn spacing = FOV Dimensioncolumn dimension / Columns
If the Field of View Shape (0018,1147) equals ROUND or HEXAGONAL:
Imager Pixel Spacingrow spacing = FOV Dimensiondiameter / Rows
Imager Pixel Spacingcolumn spacing = FOV Dimensiondiameter / Columns
In the case where the Field of View does not have the same size as the stored Pixel Data (7FE0,0010), the stored Pixel Data (7FE0,0010) may be the result of cropping, rotation, resizing and/or padding of the Field of View. The following Attributes specify the relationship of the Pixel Data area to the Field Of View Area:
In order to make use of the Pixel Data Area Origin Relative to FOV (0018,7036), the Imager Pixel Spacing (0018,1164) is needed.
| DICOM PS3.3 2026c - Information Object Definitions |
|---|