| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8-123 specifies the Attributes of the CT Reconstruction Macro.
Table C.8-123. CT Reconstruction Macro Attributes
|
Contains the Attributes holding information about the reconstruction techniques used. |
|||
|
Description of the algorithm used when reconstructing the image from the data acquired during the acquisition process. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
A label describing the convolution kernel or algorithm used to reconstruct the data. A single Value shall be present. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
A label describing the group to which Convolution Kernel (0018,1210). Required if Convolution Kernel (0018,1210) is present. May be present otherwise. |
|||
|
The diameter in mm of the region from which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region. See Section C.8.15.3.6.1. The diameter defines a circular region that is entirely contained within the encoded Pixel Data (7FE0,0010), unless the encoded image has been cropped after reconstruction. NoteIf not cropped or padded, for square images with square pixels, both Values of Pixel Spacing (0028,0030) will be equal and equal to Reconstruction Diameter (0018,1100) / Rows (0028,0010) and Reconstruction Diameter (0018,1100) / Columns (0028,0011). Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL and Reconstruction Field of View (0018,9317) is not present. Otherwise may be present if Frame Type (0008,9007) Value 1 of this Frame is DERIVED and Reconstruction Field of View (0018,9317) is not present. |
|||
|
The field of view width followed by height as used for reconstruction in mm. NoteIf not cropped or padded, for both Pixel Spacing (0028,0030) and Reconstruction Pixel Spacing (0018,9322): Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL and Reconstruction Diameter (0018,1100) is not present. Otherwise may be present if Frame Type (0008,9007) Value 1 of this Frame is DERIVED and Reconstruction Diameter (0018,1100) is not present. |
|||
|
Physical distance in the patient between the center of each reconstructed pixel, specified by a numeric pair - adjacent row spacing (delimiter) adjacent column spacing in mm. See Section 10.7.1.3 for further explanation of the order of the Values. NoteThe Values of Reconstruction Pixel Spacing (0018,9322) will equal the corresponding Values of Pixel Spacing (0028,0030) unless the pixel data has been post-processed to change the pixel size after reconstruction and before storage. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
Angle (in degrees) over which the data from which the Frame was reconstructed was collected, where 360 degrees signifies a complete revolution of the source around the gantry orbit. It is possible, in the case of over-scanning that the Reconstruction Angle is greater than 360 degrees. Shall be 0 if Acquisition Type (0018,9302) is CONSTANT_ANGLE. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
A label describing the filter applied to the reconstructed image after the original reconstruction has been completed. NoteWhen Frame Type (0008,9007) Value 1 of this Frame is DERIVED and Frame Type (0008,9007) Value 4 is FILTERED the type of filtration is described in Derivation Image Macro. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. |
| DICOM PS3.3 2026c - Information Object Definitions |
|---|