| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8-121 specifies the Attributes of the CT Position Macro.
Table C.8-121. CT Position Macro Attributes
|
Relative longitudinal position of acquisition location of this Frame in mm from an implementation specific reference point. Shall be relative to the same reference point for all Frames in this SOP Instance, but may be different from the reference point in other SOP Instances. Positions as the table (or other patient support) moves into the gantry (or as the gantry moves toward the patient support) viewed from the front are more negative. NoteRequired if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
The x, y, and z coordinates (in the Patient-Based Coordinate System) in mm of the center of the region in which data were collected. See Section C.8.15.3.6.1. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
The x, y, and z coordinates (in the Patient-Based Coordinate System) of the reconstruction center target point as used for reconstruction in mm. See Section C.8.15.3.6.1. NoteIf the reconstructed image is not magnified or panned the Value corresponds with the Data Collection Center (0018,9313). Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
| DICOM PS3.3 2026d - Information Object Definitions |
|---|