| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8-120 specifies the Attributes of the CT Table Dynamics Macro.
Table C.8-120. CT Table Dynamics Macro Attributes
|
Contains the Attributes defining the movement of the CT table. |
|||
|
The distance in mm that the table moves in one second during the gathering of data that resulted in this Frame. Table motion is relative to the gantry Frame of Reference, thus if the gantry is moving, the distance value represents the net motion. This Attribute also applies to patient support equipment other than tables. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL and Acquisition Type (0018,9302) is SPIRAL or CONSTANT_ANGLE. May be present otherwise if Frame Type (0008,9007) Value 1 of this Frame is DERIVED and Acquisition Type (0018,9302) is SPIRAL or CONSTANT_ANGLE. |
|||
|
Motion of the table (in mm) during a complete revolution of the source around the gantry orbit. Table motion is relative to the gantry Frame of Reference, thus if the gantry is moving, the feed value represents the net motion. This Attribute also applies to patient support equipment other than tables. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL and Acquisition Type (0018,9302) is SPIRAL. May be present otherwise if Frame Type (0008,9007) Value 1 of this Frame is DERIVED and Acquisition Type (0018,9302) is SPIRAL. |
|||
|
Ratio of the Table Feed per Rotation (0018,9310) to the Total Collimation Width (0018,9307). See Section C.8.15.3.4.1 for further explanation and some examples. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL and Acquisition Type (0018,9302) is SPIRAL. May be present otherwise if Frame Type (0008,9007) Value 1 of this Frame is DERIVED and Acquisition Type (0018,9302) is SPIRAL. |
The formula for Spiral Pitch Factor (0018,9311) in terms of Table Feed per Rotation (0018,9310) and Total Collimation Width (0018,9307) is:
Spiral Pitch Factor = (Table Feed per Rotation (mm))/(Total Collimation Width (mm))
An example calculation of Spiral Pitch Factor (0018,9311) for a single slice spiral acquisition of an image with a Total Collimation Width of 2.5mm and a Table Feed per Rotation of 10mm is:
Spiral Pitch Factor = (10 mm) / (2.5 mm) = 4.0
An example calculation of Spiral Pitch Factor (0018,9311) for a multiple slice spiral acquisition having a Total Collimation Width of 20mm and a Table Feed per Rotation of 10mm is:
| DICOM PS3.3 2026c - Information Object Definitions |
|---|