| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8-119 specifies the Attributes of the CT Acquisition Details Macro.
Table C.8-119. CT Acquisition Details Macro Attributes
|
Contains the Attributes defining the details of the acquisition. If Multi-energy CT Acquisition (0018,9361) is NO or is absent, only a single Item shall be included in this Sequence. If Multi-energy CT Acquisition (0018,9361) is YES, one or more Items shall be included in this Sequence. |
|||
|
References the X-Ray Path Index (0018,937A) in the Multi-energy CT Path Sequence (0018,9379) for this exposure. |
|||
|
Direction of rotation of the source about the gantry, as viewed while facing the gantry where the table enters the gantry. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL, and Acquisition Type (0018,9302) is other than CONSTANT_ANGLE. Otherwise may be present if Frame Type (0008,9007) Value 1 of this Frame is DERIVED or Image Type (0008,0008) Value 1 is DERIVED, and Acquisition Type (0018,9302) is other than CONSTANT_ANGLE. |
|||
|
The time in seconds of a complete revolution of the source around the gantry orbit. This Value is independent of the Reconstruction Angle (0018,9319) of the Frame. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL, and Acquisition Type (0018,9302) is other than CONSTANT_ANGLE. Otherwise may be present if Frame Type (0008,9007) Value 1 of this Frame is DERIVED or Image Type (0008,0008) Value 1 is DERIVED, and Acquisition Type (0018,9302) is other than CONSTANT_ANGLE. |
|||
|
The width of a single row of acquired data (in mm). NoteAdjacent physical detector rows may have been combined to form a single effective acquisition row. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise. |
|||
|
The width of the total collimation (in mm) over the area of active X-Ray detection. NoteThis will be equal to the number of effective detector rows multiplied by single collimation width. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise. |
|||
|
The distance in mm from the top of the patient table to the center of rotation of the source (i.e., the data collection center or isocenter). The distance is positive when the table is below the data collection center. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise. |
|||
|
Nominal angle of tilt in degrees of the scanning gantry. Not intended for mathematical computations. Zero degrees means the gantry is not tilted, negative degrees are when the top of the gantry is tilted away from where the table enters the gantry. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise. |
|||
|
The diameter in mm of the region over which data were collected. See Section C.8.15.3.6.1. NoteIn the case of an Acquisition Type (0018,9302) of CONSTANT_ANGLE, the diameter is that in a plane normal to the central ray of the diverging X-Ray beam as it passes through the data collection center. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL or Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise. |
| DICOM PS3.3 2026d - Information Object Definitions |
|---|