DICOM PS3.3 2022c - Information Object Definitions |
---|
Table C.8-126 specifies the Attributes of the CT Pixel Value Transformation Functional Group Macro.
This Macro is equivalent to the Modality LUT transformation in non Multi-frame IODs.
This in effect specializes the Section C.7.6.16.2.9 Pixel Value Transformation Macro.
Table C.8-126. CT Pixel Value Transformation Macro Attributes
Contains the Attributes involved in the transformation of stored pixel values. |
|||
The value b in relationship between stored values (SV) and the output units. |
|||
m in the equation specified by Rescale Intercept (0028,1052). |
|||
Specifies the output units of Rescale Slope (0028,1053) and Rescale Intercept (0028,1052). See Section C.11.1.1.2 for further explanation. If Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and Frame Type (0008,9007) Value 3 is not LOCALIZER, the value shall be HU (Hounsfield Units). |
DICOM PS3.3 2022c - Information Object Definitions |
---|