DICOM PS3.3 2026d - Information Object Definitions

C.8.13 Enhanced MR Image

This section describes the specific Modules for the Enhanced MR Image IOD.

C.8.13.1 Enhanced MR Image Module

Table C.8-79 specifies the Attributes of the Enhanced MR Image Module.

Table C.8-79. Enhanced MR Image Module Attributes

Attribute Name

Tag

Type

Attribute Description

Include Table C.8-83 “MR Image and Spectroscopy Instance Macro Attributes”

Image Type

(0008,0008)

1

Image characteristics. See Section C.8.16.1 and Section C.8.13.1.1.1.

Include Table C.8-131 “Common CT/MR and Photoacoustic Image Description Macro Attributes”

Include Table C.8-84 “MR Image Description Macro Attributes”

Samples per Pixel

(0028,0002)

1

Number of samples (planes) in this image. For Enumerated Values See Section C.8.13.1.1.2.

Photometric Interpretation

(0028,0004)

1

Specifies the intended interpretation of the pixel data. Enumerated Values are specified in the IOD that invokes this Module. See Section C.7.6.3.1.2 for definition of this term.

Bits Allocated

(0028,0100)

1

Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. For Enumerated Values See Section C.8.13.1.1.2.

Bits Stored

(0028,0101)

1

Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. For Enumerated Values See Section C.8.13.1.1.2.

High Bit

(0028,0102)

1

Most significant bit for pixel sample data. Each sample shall have the same high bit. Shall be one less than the Value in Bits Stored (0028,0101).

Pixel Representation

(0028,0103)

1

Data representation of the pixel samples. Each sample shall have the same pixel representation. For Enumerated Values See Section C.8.13.1.1.2

Planar Configuration

(0028,0006)

1C

Indicates whether the pixel data are encoded color-by-plane or color-by-pixel. Required if Samples per Pixel (0028,0002) has a Value greater than 1. See Section C.7.6.3.1.3 and Section C.8.13.1.1.2 for further explanation.

Burned In Annotation

(0028,0301)

1C

Indicates whether or not the image contains sufficient burned in annotation to identify the patient and date the image was acquired.

Enumerated Values:

NO

This means that images that contain this Module shall not contain such burned in annotations.

Required if SOP Class UID (0008,0016) is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted MR Image Storage), may be present otherwise.

Recognizable Visual Features

(0028,0302)

3

Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient.

Enumerated Values:

YES

NO

If this Attribute is absent, then the image may or may not contain recognizable visual features.

Lossy Image Compression

(0028,2110)

1C

Specifies whether an Image has undergone lossy compression (at a point in its lifetime).

Enumerated Values:

00

Image has not been subjected to lossy compression.

01

Image has been subjected to lossy compression.

Once this Attribute has been set to a Value of "01" it shall not be reset.

See Section C.7.6.1.1.5.

Required if SOP Class UID (0008,0016) is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted MR Image Storage). May be present otherwise.

Lossy Image Compression Ratio

(0028,2112)

1C

Describes the approximate lossy compression ratio(s) that have been applied to this image.

See Section C.7.6.1.1.5.2.

Required if Lossy Image Compression (0028,2110) is "01".

Lossy Image Compression Method

(0028,2114)

1C

A label for the lossy compression method(s) that have been applied to this image.

See Section C.7.6.1.1.5.1.

Required if Lossy Image Compression (0028,2110) is "01".

Presentation LUT Shape

(2050,0020)

1C

Specifies an identity transformation for the Presentation LUT, such that the output of all grayscale transformations defined in the IOD containing this Module are defined to be P-Values.

Enumerated Values:

IDENTITY

output is in P-Values.

Required if Photometric Interpretation (0028,0004) is MONOCHROME2.

Icon Image Sequence

(0088,0200)

3

This icon image is representative of the Image.

Only a single Item is permitted in this Sequence.

>Include Table C.7-11b “Image Pixel Macro Attributes”

See Section C.7.6.1.1.6.

Include Table 10-25 “Optional View and Slice Progression Direction Macro Attributes”

Include Table C.36.2.4.12-1 “RT Equipment Mapping and Plan Reference Macro Attributes”


C.8.13.1.1 Enhanced MR Image Module Attribute Description
C.8.13.1.1.1 Image Type and Frame Type

In addition to the requirements specified in Section C.8.16.1 Image Type and Frame Type, the following additional requirements and Defined Terms are specified.

These requirements and Defined Terms are also applicable to Frame Type (0008,9007).

C.8.13.1.1.1.1 Pixel Data Characteristics

Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.1. No additional requirements or Defined Terms.

C.8.13.1.1.1.2 Patient Examination Characteristics

Value 2 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.2. No additional requirements or Defined Terms.

C.8.13.1.1.1.3 Image Flavor

Table C.8-80 specifies the Defined Terms for MR additional to those defined in Section C.8.16.1.3 for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007).

Table C.8-80. MR-Specific Image Type and Frame Type Value 3

Defined Term Name

Defined Term Description

ANGIO_TIME

Angio time acquisition (peripheral vascular/carotid).

ASL

Arterial Spin Labeling.

CINE

Cardiac CINE.

DIFFUSION

Collected to show diffusion effects.

DIXON

Dixon Water Fat Imaging Techniques.

FLOW_ENCODED

Flow Encoded.

FLUID_ATTENUATED

Fluid Attenuated T2 weighted.

FMRI

Collected for functional imaging calculations.

MAX_IP

Maximum Intensity Projection.

MIN_IP

Minimum Intensity Projection.

M_MODE

Image line over time.

METABOLITE_MAP

Metabolite Maps from spectroscopy data.

MULTIECHO

Multiple echoes with different contrast weighting (e.g., proton density and T2 weighted).

PROTON_DENSITY

Proton density weighted.

REALTIME

Real-time collection of single slices.

STIR

Short Tau Inversion Recovery.

TAGGING

Images with superposition of thin saturation bands.

TEMPERATURE

Images record temperature.

T1

T1 weighted.

T2

T2 weighted.

T2_STAR

T2* weighted.

TOF

Time Of Flight weighted.

VELOCITY

Velocity encoded.


C.8.13.1.1.1.4 Derived Pixel Contrast

Table C.8-81 specifies the Defined Terms for MR additional to those defined in Section C.8.16.1.4 for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007).

Table C.8-81. MR-Specific Image Type and Frame Type Value 4

Defined Term Name

Defined Term Description


Note

This table formerly contained defined terms such as ADC, etc. These have been replaced with the use of QUANTITY and the appropriate coded entry used in Quantity Definition Sequence (0040,9220) of the Real World Value Mapping Macro. For former defined terms, see PS3.3-2018c.

C.8.13.1.1.2 Photometric Interpretation, Pixel Representation, Samples per Pixel, Planar Configuration, Bits Allocated and Bits Stored

Table C.8-82 specifies the Enumerated Values and allowed combinations of Samples per Pixel (0028,0002), Planar Configuration (0028,0006), Pixel Representation (0028,0103), Bits Allocated (0028,0100) and Bits Stored (0028,0101) for each allowable Photometric Interpretation allowed by the IOD that invokes this Module.

Table C.8-82. Allowed Combinations of Attribute Values for Photometric Interpretation, Samples per Pixel, Planar Configuration, Pixel Representation, Bits Allocated and Bits Stored

Photometric Interpretation

Samples per Pixel

Planar Configuration

Pixel Representation

Bits Allocated

Bits Stored

MONOCHROME2

1

-

0 or 1

8

8

MONOCHROME2

1

-

0 or 1

16

12, 16

RGB

3

0

0

8

8

YBR_ICT

3

0

0

8

8

YBR_RCT

3

0

0

8

8

YBR_PARTIAL_420

3

0

0

8

8

YBR_FULL_422

3

0

0

8

8

YBR_FULL

3

0

0

8

8


C.8.13.1.1.3 Pixel Presentation

See Section C.8.16.2.1.1. No additional requirements or Defined Terms.

C.8.13.1.1.3.1 Supplemental Palette Color LUTs

See Section C.8.16.2.1.1.1.

C.8.13.1.1.4 Volumetric Properties

See Section C.8.16.2.1.2. No additional requirements or Defined Terms.

C.8.13.1.1.5 Volume Based Calculation Technique

See Section C.8.16.2.1.3. No additional requirements or Defined Terms.

DICOM PS3.3 2026d - Information Object Definitions