| DICOM PS3.3 2026c - Information Object Definitions |
|---|
Table C.8-96 specifies the Attributes of the MR Diffusion Macro.
Table C.8-96. MR Diffusion Macro Attributes
|
Diffusion sensitization factor in sec/mm2. This is the actual b-value for original Frames and those derived from Frames with the same b-value, or the most representative b-value when derived from images with different b-values. Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
Specifies whether diffusion conditions for the Frame are directional, or isotropic with respect to direction. Defined Terms: Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
Sequence containing orientations of all diffusion sensitization gradients that were applied during the acquisition of this Frame. Only a single Item shall be included in this Sequence. Required if Diffusion Directionality (0018,9075) equals DIRECTIONAL May be present if Diffusion Directionality (0018,9075) equals BMATRIX. |
|||
|
The direction cosines of the diffusion gradient vector with respect to the patient Required if Frame Type (0008,9007) Value 1 of this Frame is ORIGINAL. May be present otherwise. |
|||
|
The directional diffusion sensitization expressed as a 3x3 matrix with diagonal symmetry (with six unique elements from which the other elements can be derived). The rows and columns of the matrix are the X (right to left), Y (anterior to posterior) and Z (foot to head) patient-relative orthogonal axes as defined in Section C.7.6.2.1.1. The values are in units of ms/mm2. Only a single Item shall be included in this Sequence. Required if Diffusion Directionality (0018,9075) equals BMATRIX. |
|||
|
Class of diffusion anisotropy calculation. Required if Frame Type (0008,9007) Value 4 equals DIFFUSION_ANISO. |
| DICOM PS3.3 2026c - Information Object Definitions |
|---|