| DICOM PS3.3 2026b - Information Object Definitions |
|---|
Table C.8-84 specifies the Attributes of the MR Image Description Macro.
Table C.8-84. MR Image Description Macro Attributes
|
Representation of complex data of Frames in the SOP Instance. See Section C.8.13.3.1.1 for a description and Defined Terms. Required if SOP Class UID (0008,0016) is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted MR Image Storage). May be present otherwise. |
|||
|
Indication of acquisition contrast used with Frames in the SOP Instance. See Section C.8.13.3.1.2 for a description and Defined Terms. Required if SOP Class UID (0008,0016) is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted MR Image Storage). May be present otherwise. |
|||
|
Specifies if the acquisition used and stored settling phases. |
The Value of Complex Image Component (0008,9208) shall be used to indicate which component of the complex representation of the signal is represented in the pixel data.
Table C.8-85 specifies the Defined Terms for Complex Image Component (0008,9208).
Table C.8-85. Complex Image Component Attribute Values
|
Used only as a Value in Complex Image Component (0008,9208) in the Enhanced MR Image Module if Frames within the image SOP Instance contain different Values for the Complex Image Component Attribute in the MR Frame Type Functional Group. |
Table C.8-86 specifies the Defined Terms for Acquisition Contrast (0008,9209).
Table C.8-86. Acquisition Contrast Values
|
Value should be UNKNOWN if acquisition contrasts were combined resulting in an unknown contrast. Also this Value should be used when the contrast is not known. |
|
|
Used only as a Value in Acquisition Contrast (0008,9209) in the Enhanced MR Image Module Attribute Description if Frames within the image SOP Instance contain different Values for the Acquisition Contrast Attribute in the MR Frame Type Functional Group. |
| DICOM PS3.3 2026b - Information Object Definitions |
|---|