| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table C.8.12.5-1 specifies the Attributes of the optical paths used in the VL imaging. Each optical path is a combination of illumination, filters, lenses, and sensors, and each combination is identified for possible reference by Attributes in other Modules. Additional optical path parameters may be specified in the Acquisition Context Module.
Table C.8.12.5-1. Optical Path Module Attributes
|
Number of Items in the Optical Path Sequence (0048,0105). Required if Dimension Organization Type (0020,9311) is present with a Value of TILED_FULL. May be present otherwise. |
|||
|
Describes the optical paths used during the acquisition of this image. |
|||
|
Identifier for the optical path specified in the Sequence Item. The identifier shall be unique for each Item within the Optical Path Sequence. |
|||
|
Description of the optical path specified in the Sequence Item. |
|||
|
Nominal wavelength of the illuminator in nm. Required if Illumination Color Code Sequence (0048,0108) is not present. May be present otherwise. |
|||
|
Only a single Item shall be included in this Sequence. Required if Illumination Wave Length (0022,0055) is not present. May be present otherwise. |
|||
|
Coded value for illumination method. See Section C.8.12.5.1.2. |
|||
|
Filters used in the light source (excitation) path. One or more Items are permitted in this Sequence. See Section C.8.12.5.1.3. |
|||
|
Nominal pass-through wavelength of light path filter(s) in nm. |
|||
|
Pass band of light path filter(s) in nm. This Attribute has two Values. The first Value is the shorter and the second Value the longer wavelength relative to the peak. The Values are for the - 3dB nominal (1/2 of peak) pass through intensity. One of the two Values may be zero length, in which case it is a cutoff filter. |
|||
|
Describes stack of filters used in image (emission) path between the imaging target and the optical sensor. One or more Items are permitted in this Sequence. See Section C.8.12.5.1.3. |
|||
|
Nominal pass-through wavelength of image path filter(s) in nm. |
|||
|
Pass band of image path filter(s) in nm. This Attribute has two Values. The first Value is the shorter and the second Value the longer wavelength relative to the peak. The Values are for the - 3dB nominal (1/2 of peak) pass through intensity. One of the two Values may be zero length, in which case it is a cutoff filter. |
|||
|
Nominal power of the objective lens. If a single lens is used as both condenser and objective, its power is specified in this Attribute. |
|||
|
Describes the light color sensed for each channel to generate the image. Required if this differs from the natural interpretation. Note
Shall have the same number of Items as the Value of Samples per Pixel Used (0028,0003) if present, or otherwise the Value of Samples per Pixel (0028,0002). The channels shall be described in the order in which the channels are encoded. |
|||
|
For MONOCHROME2 images, provides a transform of the pixel values with this optical path into an RGB color representation. The LUT provides a mapping to an appropriate color or pseudo-color for display purposes. |
|||
|
>>Include Table C.7-22a “Palette Color Lookup Table Macro Attributes” |
|||
|
An ICC Profile encoding the transformation of device-dependent stored pixel values into PCS-Values. See Section C.8.12.5.1.4 Required if Photometric Interpretation (0028,0004) is not MONOCHROME2, or if Palette Color Lookup Table Sequence (0048,0120) is present. NoteThe ICC Profile (0028,2000) in the Image Pixel Description Macro is not used when the Optical Path Module is used. |
|||
|
A label that identifies the well-known color space of the image. Shall be consistent with any ICC Profile (0028,2000) that is also present. See Section C.11.15.1.2. |
|||
The Optical Path Sequence (0048,0105) shall include an Item for every optical path used in the acquisition of the current image.
For Single-frame Images that do not include the Multi-frame Functional Groups Module, this Sequence shall include only one Item.
For images that use the Multi-frame Functional Groups Module, the optical path associated with each Frame is identified by reference to the Optical Path Identifier (0048,0106) of this Sequence (see Optical Path Identification Sequence (0048,0207) in Section C.8.12.6.2). The Optical Path Sequence may include Items for defined optical paths that are not referenced by any Frame.
Previously, the use of optical paths in LOCALIZER images and images referenced therein from the Section C.8.12.7 was described, but this usage has been retired. See PS3.3-2021c.
The Optical Path Identifier is not necessarily numeric; it is a string of up to 16 characters (VR SH).
Illumination Type Code Sequence (0022,0016) specifies the type of illumination of the imaging target. This Attribute may include multiple Items, as some techniques may involve multiple methods, or some method identifiers may not be completely descriptive.
For example, this Sequence might include two Items, (111741, DCM, "Transmission illumination") and (111748, DCM, "Differential interference contrast") to indicate "Transmission DIC".
The type of illumination often prescribes the set of optical path components such as filters and prisms, and their arrangement in the optical path.
For instance, phase contrast illumination will utilize a condenser annulus and a phase contrast plate; this may be indicated by an Item (111747, DCM, "Phase contrast illumination") in this Attribute, but with no further specification in the filters and lenses identification Attributes.
In some optical imaging methods, e.g., epifluorescence, there may be a single light path with its associated imaging components that is used in both illumination/excitation of the imaging target and in the transmitted/emitted path to the imaging sensor. In this case, the optical components are to be specified in the Image Path Filter Type Stack Code Sequence (0022,0018).
ICC Profile (0028,2000) provides the transform for mapping stored color pixel values into ICC standard color Profile Connection Space (PCS). For MONOCHROME2 pixels that have a preferred color rendering through the Palette Color Lookup Table Sequence (0048,0120), this Attribute provides a mapping of that preferred RGB color rendering into PCS. The ICC Profile shall follow the requirements specified in Section C.11.15.1.1.
| DICOM PS3.3 2026d - Information Object Definitions |
|---|